I.A.C.E.R. S.r.l Via S.Pertini 24/A – 30030 Martellago (Ve), Italy declares under its own responsibility that the product I-TECH PHYSIO 4 I-TECH PHYSIO EMG UMDNS Code: 11503 e 11474 (respectively) has been designed and manufactured according to the European Medical Device Directive 93/4/EEC (transposed in Italy by the D.Lgs.
Ambulatory and hospital The I-TECH PHYSIO electrotherapy medical device is available in 2 versions: I-TECH PHYSIO 4 and I-TECH PHYSIO EMG. The two devices differ only for the presence of the electromyographic module in the EMG version, by which it is possible to obtain an indication of the pre, during and post therapy of muscular activity.
The CE0068 marking on the device applies exclusively to medical programs (see the following paragraphs concerning the detailed description of the programs). Technical features Characteristics Specifications Power supply Rechargeable batteries NiMH 7.2V, 2000mAh (dimensions 6xAA) Battery charger Model EA1018G-12V 100-240VAC, 50-60Hz, 1.0A input 12VDC, 2.0A output Not replaceable 250VAC, 2A fuse Max.
Page 8
Characteristics Specifications incontinence, pains). n. 17 prog. EMS (strengthening, heating, muscular cool-down and beauty treatments). n. 18 prog. Customisable. Max out 160Vpp with a load of 1000 Ohm Frequency Between 0.5 and 40kHz Impulse Between 20 microseconds and 400 milliseconds Duration of therapy Preset programs, free memories of up to 60 minutes...
Labels Label on the back of the device Model I-TECH PHSYIO EMG Model I-TECH PHYSIO 4 Simbolo Significato Manufacturer’s logo. Product CE certification released by Notified Body N° 0068. Manufacturer. Manufacturing date (YYYY-MM). Read instructions for use. RAEE directive for the disposal of electronic waste.
N° 2 set of rectangular single-patient electrotherapy electrodes; N° 2 set of square single patient electrotherapy electrodes; • • N° 1 set of round single-patient EMG electrodes (only for I-TECH PHYSIO EMG); • N°1 set of square button single patient electrotherapy electrodes;...
The stimulation intensity is shown for each channel, on a scale of 0 to 120mA, on the I-TECH PHYSIO 4/EMG. For Sport and Beauty programs, the intensity differs according to the type of muscle or the program in use. The following description helps the user to choose the correct intensity according to impulse.
(phlebitis, thrombophlebitis, etc.). It is forbidden to use I-TECH PHYSIO4 and PHYSIO EMG if the patient has a pacemaker, is cardiopathic, suffers from epilepsy, is a pregnant woman, is an anxious person, has severe disease, has inguinal or abdominal hernias. Do not use the device if the source of the pain is unknown or not diagnosed.
Use only single-patient electrodes, supplied exclusively by the manufacturer, and take care to avoid the exchange of electrodes between different users. I-TECH PHYSIO4 and I-TECH PHYSIO EMG has been tested and guaranteed for the use only with the electrodes supplied by the manufacturer;...
Page 15
same from any responsibility related to damage to the equipment or user and will expose the user to risks such as short circuits and fire. It is forbidden: • to use the device in the presence of patient monitoring equipment, of electrosurgical (possible bruises and burns) or shortwave or microwave therapy equipment or other equipment that sends electrical impulses into the body and in general in combination to other medical devices, since it could cause...
Page 16
of the carotid sinuses; to position the electrodes near genitals and in those areas that have poor sensibility; to simulate the thyroid or apply stimulation on the neck and • mouth, as this stimulation could case important muscle spasms that can obstruct the airways, creating difficulty in breathing and problems with the heart rhythm and blood pressure;...
IF YOU HAVE ANY DOUBTS REGARDING THE DEVICE USE CONSULT YOUR DOCTOR. Patient preparation Before using I-TECH PHYSIO 4/EMG clean the skin of the area to be treated; with the cable disconnected from I-TECH PHYSIO 4/EMG, connect the electrostimulation cable jacks to the self-adhesive electrodes; position the self-adhesive electrodes on the skin (see photos of electrode positions in the Positions manual);...
Device use Instructions on use for a single patient Turn on I-TECH PHYSIO 4/EMG by pressing ON/OFF button on the small rear panel of the device. Select single mode and press the OK key to confirm.
Page 19
This enables the user to know exactly when a contraction starts. At the end of the first phase, I-TECH PHYSIO 4/EMG At the end of the first phase, the I-TECH PHYSIO 4/EMG cancels the previously selected intensity, and warns the user with an intermittent signal;...
Press the OK button twice consecutively to stop the program. Instructions on use for two patients Turn on I-TECH PHYSIO 4/EMG by pressing the ON/OFF button on the small rear panel of the device. Select the two patient mode and press the OK key to confirm.
Page 21
local anaesthetic effect, TENS burst which is a mix of the first two types of TENS. The rehabilitative action of TENS is represented by its power to reduce pain thereby restoring physiological conditions; most of the time this allows the patient to regain normal motor function. Consider a patient suffering from irritating periarthritis;...
Page 22
the stimulation does not become more effective, just more irritating, so it is best to stop before that point. Specific programs Medical Progr Prog. Description PHASE 1 Yes/NO Total time 30min Rapid TENS Frequency 90Hz Impulse width 50µS Total time 20min Endorphinic TENS Frequency 0,5Hz 0,5Hz...
Page 23
Medical Progr Prog. Description PHASE 1 Yes/NO Frequency 1500Hz Impulse width 100µs (intensity limited to 50 steps) Total time 20min Haematomas Pulses for hematomas Oedema Total time 15min Diadynamic 1 (MF) Total time 15min Diadynamic 2 Total time 15min (MFSR) Frequency 2s ON, 1s OFF Diadynamic 3 Total time 15min...
Page 24
impulses leave part of the body (for example the hand) and run along the nerve tracts (through small-diameter nerve fibers) until they reach the central nervous system where the impulses are interpreted as pain. Conventional TENS activates large-diameter nerve fibers, blocking the path of small-diameter nerve fibers at the spine.
Page 25
W03 • Endorphinic TENS 1 Hz Frequency (medical program) Cable type: TYPE 1 (see Figure 1 - Cable type.) Like the previous one with 1Hz frequency. W04 • Endorphinic TENS 2 Hz Frequency (medical program) Cable type: TYPE 1 (see Figure 1 - Cable type.) Like the previous one with 2Hz frequency.
Page 26
Intensity: it is the maximum tolerable value (well in excess of conventional tens, and therefore with considerable contraction of the muscles surrounding the area treated). W07 • TENS Burst impulse (medical program) Cable type: TYPE 1 (see Figure 1 - Cable type.) This program produces a training effect using the frequencies of conventional TENS.
Page 27
Intensity set on the Effective display current 0,90 mA 1,5 mA 2,7 mA 3,6 mA Ionophoresis For the ionophoresis programs the stimulation intensity shall be adjusted so as to feel a remarkable tingling in the treated area, producing a slight contraction of the surrounding muscles. If you feel discomfort (or pain), reduce the intensity and eventually stop the therapy.
Page 28
Table of the main drugs used in the iontophoresis treatments Drug Polarity Prevalent action Indications Sclerolytic, Scars, Potassium iodide Negative emollient Dupuytren's disease, keloids Acetylsalicylate Negative Analgesic Arthrosis lysine Extra / intra articular Flectadol, Aspegic Negative Analgesic arthrosis, rheumatism Local anesthetics Local anesthesia, (novocaine, Negative...
Page 29
(channel 1 or channel 2, choose both channels only if you want to stimulate two different areas) and turn I-TECH PHYSIO 4/EMG on. If you wish to double the number of electrodes, you can request an additional ionophoresis kit by contacting directly the Figure 3 –...
Page 30
Contact the manufacturer for other information. W09-W10 • Iontophoresis 1/Iontophoresis 2 (medical programs) Cable type: TYPE 1 (see Figure 1 - Cable type.) At the end of the program, the skin could turn slightly red; any reddening usually vanishes a few minutes after the end of the program. Select channel 1 or channel 2 (select both channels only if you need to stimulate two different areas).
Page 31
W12 • Oedema (medical program) Cable type: TYPE 1 (see Figure 1 - Cable type.) Consult a doctor before using this program to treat oedemas. A combination of various types of square-wave impulses has a graduated draining effect on the area to be treated (impulses at different frequencies drain the area at different depths).
Page 32
W14 • MFSR Diadynamic – Mono-Phase Quick Syncopation (medical program) Cable type: TYPE 1 (see Figure 1 - Cable type.) It is a sinusoidal single-phase current with pulse trains of 10ms and a 10ms pause, with 2 seconds of action and a 1 second pause at 50 Hz. The main effect of this kind of current is an excitomotor action on striated muscle and on motor nerves, but it also has an analgesic, tonic and vasodilatory effect.
Page 33
vasodilatation, has a sedative effect on the sympathetic nervous system and a great effect of functional and motor recovery of the muscle. Indication: it is indicated for the therapy of painful states of non-spastic origin, to improve the tone of the connective tissue of muscles. Session duration: 15 minutes.
Page 34
W19 • CP Current – Short period (medical program) Cable type: TYPE 1 (see Figure 1 - Cable type.) It is composed of alternated single-phase (3 seconds) and biphasic (3 seconds) waveforms. The short term current has an essentially dynamogenic effect. This action promotes contraction of the striated muscles, improves the nutritional status of the tissue and facilitates the re- absorption of post-traumatic oedemas.
Page 35
the migration of vascular cells and the synthesis of collagen occur, increasing the speed of recovery. If the cathode (negative electrode) is placed over the wound, a bactericide effect occurs and the growth of pathogenic microorganisms is delayed. Session duration: 15 minutes. Electrodes’...
Page 36
any channel: phase 2 will begin automatically at the selected intensity (90% first turn, 95% second turn, 100% third turn). − The second phase lasts 15 minutes and it alternates 10 seconds of recovery with 5 seconds of contraction. W23 • Neofaradic 20Hz (non-medical program) Cable type: TYPE 1 (see Figure 1 - Cable type.) It is a low frequency (20 Hz) current that is used to produce individual muscular contractions.
W26 • Interferential (medical program) Cable type: TYPE 1 (see Figure 1 - Cable type.) interferential therapy based interference of two sinusoidal currents at different frequencies applied to the patient; the resultant current, endogenously generated, is a new kind of current.
Page 38
Medical Prog. Description PHASE 1 PHASE 2 PHASE 3 Yes/No ramp up e 3s ramp down Total time Total time 3min 10min (3Hz- Frequency Stress 150µs x 9s + incontinence 50Hz-150µs x Pulse width 6s) x 40 cycles 150µs Total time Total time 3min 10min...
Page 39
Medical Prog. Description PHASE 1 PHASE 2 PHASE 3 Yes/No Frequency Pulse width 90Hz 80µs Pulse width Burst train 60µs Total time 15min Total time Sprains/ Frequency 15min Bruises 90Hz Pulses for Pulse width hematomas 60µs Total time Total time Total time 15min 10min...
Page 40
Medical Prog. Description PHASE 1 PHASE 2 PHASE 3 Yes/No Total time 10min (3Hz- Total time 1min Total time 200µs x 7s Frequency 30min +1Hz- Periarthritis 150Hz Frequency 90Hz 200µs x Pulse width 3s+ 30Hz- Pulse width 60µ 200µs 200µs x 5s) x 40 cycles Total time...
Page 41
Medical Prog. Description PHASE 1 PHASE 2 PHASE 3 Yes/No limitation I a 3,6mA Total time Muscular 20min with tears limitation I a 3,6mA Total time 30min Frequency Herpes Zoster 70Hz Pulse width 20µS Wound/heali Total time 15min Wounds/bact Total time ericide 15min Tempo tot...
Page 42
longer duration (calculated in milliseconds and not in microseconds as in the case of normal innervated muscle) and a much lower frequency. Session duration: 20 minutes, in one phase. Electrodes’ positioning: use 2 large electrodes, better if wet and in sponge, placed at the heads of the muscle to be stimulated.
Page 43
stimulation probes, supplied separately together with the relative instructions. Consult your doctor before using this program and during the treatment period. Session duration: 13 minutes. Intensity: to be adjusted above the perception threshold to produce light internal stimulation. R05 • Mixed incontinence (medical program) Cable type: TYPE 1 (see Figure 1 - Cable type.) This program, for which no photographs are provided regarding the position of the electrodes, requires the use of appropriate internal...
Page 44
carried out on a daily basis; proceed with the treatment until the symptoms pass. Session duration: 33 minutes. Electrodes’ positioning: photo 25 of the Positions manual. Intensity: to be adjusted to a level between the thresholds of perception and pain: the maximum intensity level is the moment in which the muscles surrounding the treated area begin to contract.
Page 45
R10 • Hand and wrist pain (medical program) Cable type: TYPE 1 (see Figure 1 - Cable type.) This program is suitable for all types of hand and wrist pain (for example Carpal Tunnel Syndrome). The intensity shall be adjusted 3 times, one for each time one of the phases begins.
Page 46
R13 • Epitroclea (medical program) Cable type: TYPE 1 (see Figure 1 - Cable type.) Also known as “golfing elbow”, it affects golfers but also those who carry out repetitive tasks or tasks involving frequent intense strain (for example carrying a particularly heavy suitcase). It causes pain in the flexor and pronator tendons inserted in the epitroclea.
Page 47
R15 • Neuralgias (medical program) Cable type: TYPE 1 (see Figure 1 - Cable type.) Program suitable for people suffering from neuralgie. It is recommended 2 daily treatments for 10/12 days Session duration: 30 minutes, in one phase. Electrodes’ positioning: identified the area to be treated, position the electrodes as shown in Figure 2.
Page 48
R20 • Muscle tears (medical program) Cable type: TYPE 1 (see Figure 1 - Cable type.) active channels: CH1 Session duration: 20 minutes, in one phase. Electrodes’ positioning: 2 electrodes of channel 1, over the affected area. Intensity: adjusted between 6 and 12. The maximum selectable intensity is R21 •...
Page 49
Intensity: adjusted between the threshold of perception and the pain threshold. R24 • Venous insufficiency (medical program) Cable type: TYPE 1 (see Figure 1 - Cable type.) The program is indicated for the treatment of venous insufficiency. It is possible to use square or rectangular electrodes according to the size of the calf and the comfort of the patient.
Table of current density with electrodes measuring 48x48mm: Value displayed Current density Current density mA/cm² on screen μA/cm² 0,034 0,068 0,137 0,205 Table of current density with electrodes measuring 50x90mm: Value displayed Current density Current density mA/cm² on screen μA/cm² 0,0175 17,5 0,035...
Page 51
Medical Progr. Description PHASE 1 PHASE 2 PHASE 3 PHASE 4 Yes/No 5min 10min 10min (30rip) 5min Frequency (30rip) 14s a 4Hz Frequenc 13s a 6Hz 1r+4+1r sec y 3Hz 1r+5+1r s at 70Hz at 50Hz Total time 10min Total time Total time Total time (30rip)
Page 52
Medical Progr. Description PHASE 1 PHASE 2 PHASE 3 PHASE 4 Yes/No 4min at 8Hz, 1min at 3Hz Total time 15min, Muscle Recovery Freq. step 6/4/2Hz Total time 20min Total time Total time Altern. Contr. CH 1/2 3/4, 3min 10min Agonist/ 6s a 8Hz Antagonist...
Page 53
Medical Progr. Description PHASE 1 PHASE 2 PHASE 3 PHASE 4 Yes/No Total time 15min Frequency 12Hz Microlifting Pulse width 100µs Total time Total time Total time Total time 10min 10min 5min 5min Atrophy 9s at 4Hz 9s at 4Hz prevention Frequency Frequenc...
Page 54
Medical Progr. Description PHASE 1 PHASE 2 PHASE 3 PHASE 4 Yes/No Pulse sequence 4 channels Pulse width on 4 (40-300µS), width 300µs channels 3r+3r sec at 300µs (40- 60Hz 300µS), 3r+3r sec at 60Hz ATTENTION! Intensity of stimulation: EMS programs (Firming up and Toning up, Strength, Endurance, Agonist/Antagonist) are divided in 3 phases: warm up phase, work phase, recovery phase.
Page 56
E03 • Base strenght (non-medical program) Cable type: TYPE 1 (see Figure 1 - Cable type.) The basic strength program is used in sport to develop basic strength which, by definition, is the maximum tension that a muscle can exert against constant resistance.
Page 57
E04 • Fast strenght (non-medical program) Cable type: TYPE 1 (see Figure 1 - Cable type.) This program is designed to increase speed in fast athletes and develop it in athletes lacking the quality. The exercise assumes a fast pace and the contraction is short, as is the recovery.
Page 58
usually best to complete a three-week basic strength cycle of increasing intensity before using this program. Then continue with three weeks of explosive strength twice a week. Intensity: during contraction the intensity must be the highest that can be endured in order to obtain maximum muscle exertion whilst involving the greatest number of fibres (over 35mA).
Page 59
ATTENTION: the program involves an automatic increase in intensity between the second and the third phase. Consequently it will not be necessary to adjust stimulation intensity between phases but only press the Up arrow key for any one of the 4 channels.
Page 60
cycle will go up to 95% of intensity and finally the third work/recovery cycle will go up to 100% of the intensity set for phase 2. Session duration: 50 minutes. Electrodes’ positioning: from Position manual: Biceps (photo 02/15), Triceps (photo 03/16), Hand Extensors (photo 04), Hand Flexors (photo 05), Deltoid (photo 06).
Page 61
Session duration: 15 minutes, in one phase. Electrodes’ positioning: photos from 01 to20 of the Position manual. Intensity: medium/low(15÷25mA), with increment in the last 5 minutes. E10 • Agonist / Antagonist (non-mediacal program) Cable type: TYPE 1 (see Figure 1 - Cable type.) The electric stimulator produces alternating contractions on the 4 channels: during the first phase of warm-up the 4 channels work simultaneously, during the second work phase muscle contractions are...
Page 62
Phase 1 Phase 2 Fase 3 20min work alterned over channel pairs: 3min 6s recovery 10min Warm up 7s work on CH1/CH2 + 7s work on Recovery CH3/CH4 E11 • Lipolysis (non-medical program) Cable type: TYPE 1 (see Figure 1 - Cable type.) This program is widely used in the field of beauty treatments to increase micro-circulation and promote lymphatic activity in areas where there are fatty deposits.
Page 63
E12 • Drainage (non-medical program) Cable type: TYPE 1 (see Figure 1 - Cable type.) The purpose of these programs is to promote drainage, increasing microcirculation inside and around treated muscular fibres, also creating rhythmic contractions to facilitate the flow of algogenic substances and lymphatic activity.
Page 64
stimulation (for example 3÷10mA) and then with caution increase until you reach a good activation of the muscles. IMPORTANT: it is not necessary get to intensity levels such that provide discomfort! The equation more pain = more benefit is completely misleading and counterproductive. Great and important goals are achieved with perseverance and patience.
Page 65
Phase 1 Phase 2 Phase 3 Phase 4 Session duration Session 5min 10min duration 10min 5min Warm up 9s recovery 9s recovery Recovery 6s work 6s work E15 • Atrophy (medical program) Cable type: TYPE 1 (see Figure 1 - Cable type.) This program acts selectively on slow-twitch fibres.
Intensity: must be such to assure good muscular contractions during treatment, but not so strong as to generate soreness (20÷25mA). Session duration: 25 minutes. Electrodes’ positioning: from the Positions manual: Muscles Group all you want to stimulate in a sequence Phase 1 Phase 2 Phase 4...
Page 67
Medical Prog. Description PHASE 1 Yes/No Adjustable pulse width 20-500µs Total time from 1 to 60min TENS Spyke Adjustable frequency 0,25-120Hz Adjustable pulse width 20-500µs Total time from 1 to 60min Recovery frequency 0-12Hz, Contraction M08- frequency 20-120Hz EMS 1s slope Recovery period 2-28s, Contraction period 2-10s Pulse width 60-500µs...
Page 68
M01/M02 • Denervated 1/2 (medical programs) Cable type: TYPE 1 (see Figure 1 - Cable type.) active channels CH1 and Stimulation programs of the denervated muscle. Electrodes’ positioning: use 2 big electrodes, better if wet and in a sponge, placed at the two ends of the muscle to be treated. −...
Page 69
To move the cursor and select the stimulation parameter, use the − right and left arrow keys, change the values with the up and down arrow keys. − Once all the stimulation parameters have been set, position the cursor on "OK" to memorize the program. start stored program,...
Page 70
Table of settable values: Parameter Description Minimum Maximum Frequency 0,25 Hertz 120 Hertz Impulse width 20 microseconds 500 microseconds Program total time 1 minute 60 minutes M08/M09 • EMS ramp of 1 second (non-medical programs) Cable type: TYPE 1 (see Figure 1 - Cable type.) ATTENTION: 1 second slopes make for a faster contraction, while 2/3 second slopes make for a more progressive contraction.
Page 71
Parameter description Minimum Maximum Program total time 1 minute 60 minutes M10/M11 • EMS ramp of 2 seconds (non-medical programs) Cable type: TYPE 1 (see Figure 1 - Cable type.) Table of settable values: Parameter description Minimum Maximum Contraction frequency 20 Hertz 120 Hertz Contraction period...
Page 72
M16/M17/M18 • Agonist/antagonist (non-medical programs) Cable type: TYPE 1 (see Figure 1 - Cable type.) Available programs to set treatments with alternated contractions on a pair of channels. Electrodes’ positioning: to refer to the photos of the Position manual. − To move the cursor and select the stimulation parameter, use the right and left arrow keys, change the values with the up and down arrow keys.
To access to the functionality of the EMG module, simply click on the EMG icon in the I-TECH PHYSIO EMG main menu: the channel setup menu will be displayed with a screen similar to the one in the figure below.
Page 74
It’s possible activate or deactivate the acoustic signal of work Clean the skin near the area to be treated; connect the amplifier to I-TECH PHYSIO EMG; connect the 2 concentric electrodes to the cables with dedicated connector and the reference pregelled electrode squared to the 2mm male plug.
Page 75
(each channel is associated with a pair of concentric electrodes and a common mass) . A screen similar to the one in the figure below will appear: In this mode it is necessary to record the reference contraction level to work on.
Page 76
The distance between the dashed vertical bars corresponds to the precision required to maintain the desired level. This precision can be changed by accessing the channel setup menu. Work margin, described above. The Target value set in the TAR (Target, Keep Level) section is calculated as the average of the amplitudes of the active channels.
It is recommended that the manufacturer carries out a functional test every 24 months. The manufacturer does not consider the I-TECH PHYSIO 4/EMG device repairable by any personnel outside the company. Each...
Precaution for the storage There is no particular precaution to be taken during transportation of the device, since I-TECH PHYSIO 4/EMG is a portable device. In any case it is recommended to store I-TECH PHYSIO 4/EMG and its accessories in the supplied carrying bag after each treatment.
Use a dry cloth to clean the battery charger. Do not open the battery charger: it does not contain repairable parts. Disposal The I-TECH PHYSIO 4/EMG devices, was designed and engineered to have minimal negative environmental impact, in consideration of its performance and safety requirements, following the disposition given by the European Directive 2012/19/EU, regarding the waste of electrical and electronic equipment.
This symbol means that the product shall not be disposed as domestic waste. The correct disposal of obsolete equipment, accessories and most of all of batteries contributes in preventing possible negative consequences on human and environmental health. The user must dispose of scrap equipment by taking it to a recognized center for recycling of electrical and electronic equipment.
Page 81
2) The warranty period is valid only on the electronic parts. The warranty will be granted by the shop or directly by the manufacturer. 3) The warranty covers only the product damages, which causes its malfunctioning. 4) Warranty means that only the manufacturing defect components or material are covered by reparation or free substitution, hand work included.
Page 82
The I-TECH PHYSIO 4 and I-TECH PHYSIO EMG devices has been designed and manufactured according to the international ELECTROMAGNETIC COMPATIBILITY legislation EN 60601-1-2:2015 with the aim of providing adequate protection from harmful interference when installed in homes and health establishments.
Page 83
ALL EQUIPMENT AND SYSTEMS The I-TECH PHYSIO 4/EMG is intended for use in the electromagnetic environment specified below. The customer or the user of the I-TECH PHYSIO 4/EMG should assure that it is used in such an environment. Electromagnetic environment -...
Page 84
Guidance and manufacturer’s declaration – ELECTROMAGNETIC IMMUNITY – FOR ALL EQUIPMENT AND SYSTEMS The I-TECH PHYSIO 4/EMG is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in...
Page 85
Guidance and manufacturer’s declaration – ELECTROMAGNETIC IMMUNITY – FOR EQUIPMENT AND SYSTEMS THAT ARE NOT LIFE-SUPPORTING The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment Test level Conformity...
Page 86
I-TECH PHYSIO 4/EMG that are not life-supporting I-TECH PHYSIO 4/EMG is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of I-TECH PHYSIO 4/EMG can help prevent electromagnetic interference by maintaining a minimum distance between...
Page 87
I-TECH PHSYIO 4/EMG All rights reserved I-TECH PHYSIO 4/EMG and the logo are property exclusively of I.A.C.E.R. Srl and registered. Edition: MNPG178-05 of the August 10 , 2020 IACER Srl MNPG178...