Advertisement

Quick Links

USER MANUAL
MNPG263-01 Edition 03/05/19
Magnetotherapy model
LaMagneto Pro
I.A.C.E.R. Srl
www.itechmedicaldivision.com

Advertisement

Table of Contents
loading

Summary of Contents for I-Tech LaMagneto Pro

  • Page 1 USER MANUAL MNPG263-01 Edition 03/05/19 Magnetotherapy model LaMagneto Pro I.A.C.E.R. Srl www.itechmedicaldivision.com...
  • Page 2 I.A.C.E.R. Srl Via S. Pertini 24/A – 30030 Martellago (VE) ITALY Tel. +39 041 5401356 – Fax +39 041 5402684 e-mail: iacer@iacer.it https://www.itechmedicaldivision.com...
  • Page 3: Table Of Contents

    Summary Summary Introduction Magnetotherapy Technical specification Manufacturer Declaration of conformity Classifications Intended purpose and scope of use Technical characteristics Labelling Device labels in detail Description of the symbols (device and packaging) Contents of the pack H ow to use the device Warnings Electromagnetic interference Contraindications and side effects...
  • Page 4: Introduction

    Introduction Magnetotherapy The treatment of certain conditions through low frequency and high intensity pulsed magnetic fields has garnered great consensus amongst international scientific circles for many years, especially as regards chronic and degenerative diseases. Magnetotherapy uses low frequency and high intensity pulsed magnetic fields induced by the electric current that runs through a coil;...
  • Page 5: Technical Specification

    Notified body:TÜV Rheinland Italia srl, Via Mattei 3 - 20010 Pogliano Milanese (MI) Italy. Certification path:Annex II, excluding item 4. The LaMagneto Pro device is a class IIa equipment according to Annex IX, rule 9 of Directive 93/42/EEC (and subsequent amendments). Martellago, 13/03/18...
  • Page 6: Classifications

    LaMagneto line. Therefore the classification is IIa. The LaMagneto Pro device assumes the following classifications: • Class II with type BF applied part (Classif. IEC EN 60601-1);...
  • Page 7: Intended Purpose And Scope Of Use

    LaMagneto Pro is particularly indicated in the treatment of bone fractures even in the presence of rigid bandages or plaster casts. The distinctive feature of the LaMagneto Pro device is the possibility of managing two treatments at the same time, by setting different times and programs.
  • Page 8: Technical Characteristics

    Technical characteristics Power supply Pow. UES36LCP-150200SPA, out 15VDC-2A Max. current consumption Insulation class (CEI EN 60601-1) Applied part (CEI EN 60601-1) Dimensions (length x width x height) 180x110x50 (mm) Intensity of the field Adjustable with increasing scale up to 200 Gauss (per channel).
  • Page 9: Labelling

    Labelling Label 3 Label 2 Label 4 Label 1 (on the back) Device labels in detail Label 1 Label 2 Label 3 Label 4 DC 15V/2.0A ON/OFF CH1 CH2 UES36LCP-150200SPA I.A.C.E.R. Srl MNPG263-01...
  • Page 10: Description Of The Symbols (Device And Packaging)

    Relative humidity Manufacturer's data IP21 Degree of protection against the entry of solids, powders and liquids Contents of the pack The LaMagneto Pro pack contains: • N°1 LaMagneto Pro device; • N°1 medical power supply (approx 1.5mt cable); • N°1 Use and maintenance manual;...
  • Page 11: How To Use The Device

    • Magnet for verifying therapy operation • Non-woven fabric strip 15x150 cm (see page 15 for further details) The professional applicator with a pair of solenoids is available as an optional accessory. Visit www.itechmedicaldivision.com for more information. How to use the device Warnings It is recommended to read this manual carefully before using the device.
  • Page 12: Electromagnetic Interference

    • the electrical system of the environment in which LaMagneto Pro is inserted complies with national laws. • the devices are used in strict compliance with the instructions reported in this manual.
  • Page 13: Contraindications And Side Effects

    Quick use of the device with preset parameters To start using LaMagneto Pro quickly and easily, we recommend that you follow the steps below: 1. Connect the applicator (or applicators) to the device by connecting the plug of the applicator cable to one of the two ports (CH1-CH2) on the panel at the top of the appliance;...
  • Page 14 6. Press OK.The display will show the basic setting time of the therapy (2 hours) and magnetic field intensity. These are the average values suggested by IACER to immediately start the treatment effectively. 7. Press the OK button.The device will start the treatment, displaying the magnet icon with the magnetic field flux.The green light below the display notifies the therapy is underway.
  • Page 15: Use Of The Therapeutic Belts And Solenoids, Main Applications And Suggestions

    Note: the device recognises if the applicators are connected correctly. During the therapy phase, the connection status is displayed below the magnet icon. The presence of the symbol next to the channel number (1 or 2) confirms the applicator is connected correctly and recognised.The symbol next to the channel number (1 or 2) tells you that the applicator is not connected correctly, missing or not working...
  • Page 16 Knee Back (lumbar) Ankle Cervical vertebrae Elbow Shoulder Wrist Knee Ankle Tibia / fibula Femoral head Femur Shoulder I.A.C.E.R. Srl MNPG263-01...
  • Page 17: Instructions For Using The Preset Programs

    230VAC, 50-60 Hz); 4. Press the ON/OFF switch on the small panel in the upper part so it is in the ON position: the display will show the I-TECH MEDICAL DIVISION logo and then the main menu screen; I.A.C.E.R. Srl...
  • Page 18 5. Select "Single Patient" and scroll through the programs using the buttons and select the desired program 6. Press OK. The display will show the basic setting of therapy time (2 hours) and intensity of magnetic field (intensity 40) that we will modify;...
  • Page 19 b) Use the buttons to set the desired therapy hours (from 0 to 24) and confirm by pressing the OK button. The display will highlight the minutes of therapy; c) Use the buttons to set the desired therapy minutes (from 0 to 59) and confirm by pressing the OK button. The display will highlight the intensity of the treatment;...
  • Page 20: Instructions For Using The Free Memory Slots

    230VAC, 50-60 Hz); 4. Press the ON/OFF switch on the small panel in the upper part so it is in the ON position: the display will show the I-TECH MEDICAL DIVISION logo and then the main menu screen; 5. Select "Single Patient" and press the arrow to select "Memory...
  • Page 21 6. Press OK. The display will show the basic setting of therapy time (2 hours) and intensity of magnetic field (intensity 40) that we will modify; a) Press the button so that the spanner is highlighted. At this point press OK: the display shows a moving spanner icon on the left-hand side;...
  • Page 22: Instructions For Using Programs In Double Patient Mode

    At the end of the therapy, the device will automatically return to the program menu screen. Instructions for using programs in Double Patient mode To use the "Double Patient" mode of the LaMagneto Pro device, follow these simple steps: 1. Connect the applicator (or applicators) to the device by connecting the plug of the applicator cable to one of the two ports (CH1-CH2) on the panel at the top of the appliance;...
  • Page 23 Choose the desired settings: a) Use the buttons to set the desired therapy hours (from 0 to 24) and confirm by pressing the OK button. The display will highlight the minutes of therapy; b) Use the buttons to set the desired therapy minutes c) (from 0 to 59) and confirm by pressing the OK button.
  • Page 24 d) Use the buttons to set the treatment intensity and confirm by pressing the OK button; e) The device will now let you select the second program to be carried out. Follow the procedure described in point 6. 7. At the end of the therapy, the device will automatically return to the program menu screen.
  • Page 25: List Of Stored Programs

    List of stored programs ATTENTION: programs from 42 to 63 are programs to be used exclusively following specific medical advice. Preset values Recommended values Condition Duration Session Interval cycles between hours sessions Osteoporosis 2 – 6 30 – 60 24 hours Arthrosis 2 –...
  • Page 26 Algodystrophy 2 – 6 30 – 60 24 hours Cartilage lesion 2 – 6 30 – 60 24 hours Ligament lesion 2 – 6 30 – 60 24 hours Bone oedema 2 – 6 30 – 60 24 hours Tendinitis 2 –...
  • Page 27 Treat. 1 Hz 24 hours Free Free Treat. 3 Hz 24 hours Free Free Treat. 5 Hz 24 hours Free Free Treat. 8 Hz 24 hours Free Free Treat. 10 Hz 24 hours Free Free Treat. 15 Hz 24 hours Free Free Treat.
  • Page 28 Cervicalgias: specific program for the reduction of cervical pain. Chronic pain: program designed to reduce pain and inflammatory processes. It is recommended to use the applicators on the area where the pain is felt. Fractures: specific program for the stimulation of bone regeneration in a post-traumatic condition.
  • Page 29 Bone oedema: specific program for the stimulation of bone tissues in cases of bone oedema. Tendinitis: program designed for the treatment of inflammation of tendon tissues. Chondropathy: specific program for osteoarthritis (a particular type of chondropathy), designed for the reduction of inflammation of cartilaginous tissues.
  • Page 30 Tarsal tunnel syndrome: program designed to reduce the pain of this condition and other polyneuropathies. Programs from 47 to 63 vary in frequency but are the same in terms of waveform. In particular, programs from 42 to 45 (night) are indicated for a therapy time ranging from 6 to 10 hours, during sleep.
  • Page 31: Last 10

    Chondrites, rheumatic diseases The therapy duration values are those recommended by IACER S.r.l. and can be altered by the user. The LaMagneto Pro magnetotherapy device incorporates the indications regarding magnetic field, frequency of therapy and power delivered that are found in scientific and medical literature, the result of experiments and clinical evaluations carried out (Barker - Lunt 1983, Bassett - Pawluk - Pilla 1974, Bassett - Valdes - Hernandez 1982).
  • Page 32: Settings (Language Selection)

    Settings (language selection) Press the ON/OFF switch located on the small panel in the upper part so it is in the ON position. After the I-TECH MEDICAL DIVISION logo appears, press and select the "Settings" menu. At this point select "Language"...
  • Page 33 From the main menu press until "Lock" is selected, then press OK. Choose the therapy from those listed by pressing the buttons followed by OK. Choose the desired settings: a) Use the buttons to set the desired therapy hours (from 0 to 24) and confirm by pressing the OK button. The display will highlight the minutes of therapy;...
  • Page 34: H Ow To Lookafter

    c) Use the buttons to set the treatment intensity and confirm by pressing the OK button; d) Confirm the device lock by pressing the OK button. At this point all the functions of the device are locked except for the "Single Patient"...
  • Page 35: Cleaning The Device

    Transport and storage Transport precautions There is no particular care to be taken during transport as LaMagneto Pro is a portable device. It is recommended to store LaMagneto Pro and its accessories in the bag supplied after each use and store everything inside the original box.
  • Page 36: Disposal

    It is recommended not to twist the power supply and applicator cables. Storage precautions The storage location should have the following characteristics: ambient temperature from +5° to +40°C. relative humidity from 15 to 93% pressure from 700 to 1060 hPa Disposal The product is subject to the WEEE regulation (the symbol is present on the label...
  • Page 37: Support

    EMC Tables Emission aspects Emission test Compliance Electromagnetic environment - guidance RF emissions Group 1 The LaMagneto Pro product uses RF Cispr 11 energy only for its internal operation. Therefore, its RF emissions are very I.A.C.E.R. Srl MNPG263-01...
  • Page 38 RF emissions Class B The LaMagneto Pro is suitable for use Cispr 11 in all buildings in addition to domestic ones and those directly connected to a low-voltage power supply network that...
  • Page 39 Immunity aspects The LaMagneto Pro is intended to work in the electromagnetic environment specified below.The customer or user should make sure that it is used in such environment Immunity test Test level Compliance Electromagnetic 60601-1-2 level environment - guidance ± 8kV contact ±...
  • Page 40 Recommended separation distance between portable and mobile radio communication devices and the LaMagneto Pro device The LaMagneto Pro is intended to operate in an electromagnetic environment in which RF irradiated disturbances are controlled. The customer or the operator of the device can help...
  • Page 41 For transmitters specified for a maximum output power not listed above, the recommended separation distance d in metres (m) can be calculated using the equation applicable to the frequency of the transmitter, where P is the maximum output power of the transmitter in watts (W) according to the transmitter manufacturer.
  • Page 42: Warranty

    Warranty Make reference to the national laws for any warranty conditions by contacting the national distributor (or directly the manufacturer IACER). LaMagneto Pro. All rights reserved.LaMagneto Pro and the logo are the exclusive property of I.A.C.E.R.Srl and are registered trademarks.
  • Page 44 Unlock code: 18273...

Table of Contents