2-1. Preparations

2-1-1. Equipment Required

. Oscilloscope
Tektronix 2465 or equivalent
(band width: 350 MHz or more)
. NTSC, PAL, SECAM component signal generator
Tektronix TG2000 + AVG1 (optional module)
+AWVG1 (optional module) or equivalent
. VG (Programmable video signal generator)
VG814 or equivalent
. Digital voltmeter
Advantest TR6845 or equivalent
. Luminance meter
n
Perform the following adjustments at least 5 minutes after
turning on the power.

2-1-2. Setting the Factory Mode

1. Check that the STATUS is ON in MENU.
2. Exit MENU.
3. Press the keys in the following order;
"ENTER" → "ENTER" → "
4. The message "Do you wish to enter into the FACTO-
RY MODE?" YES : 9 NO : 0 will be displayed.
5. Select YES : 9.
6. When returning from FACTORY MODE to the USER
MODE, repeat the Step 3.
7. The message "Do you wish to return into the USER
MODE? YES : 9 NO : 0 will be displayed.
8. Select YES : 9.
VPL-FX51
Section 2

Electrical Adjustments

2-2. Highland Mode Setting

In case this model uses above sea level more than 1500
meter, set the model to the Highland Mode as follow items.
This model can be used above sea level more than from
1500 m to 2600 m.
Procedure
1. Enter the FACTORY mode.
2. Select Other of device adjust on the menu and depress
3. Select Other/Highland Mode by depressing 9 and 0
4. Set "1" with using 8 key.
5. Return the Device Adjust page by depressing the
6. Select the item SAVE TO MEMORY and save it by
n
When releasing the Highland Mode, set "0" in the item 4.

2-3. R, G, B, VCOM Adjustment

" → "ENTER".
1. Enter the P.DRV of the Device Adjust in the Factory
2. Entering the item 04 VCOM R, check 1 line ON/OFF
3. Set the fixture and Luminance meter to the Lens first
4. Adjust VCOM G and VCOM B in the same manner.
5. Save the data by selecting the item " SAVE TO
6. Reverse top and down, value of each R. G. B is
7. Memorize.
the Enter.
keys.
Enter.
depressing the Enter.
mode.
signal.
and connect the Oscilloscope. Observing output from
the Luminance meter through the Oscilloscope (AC
mode), adjust the waveform using the key 8 or 7,
to minimize flickers. (Neglect intermittent very small
distortion)
MEMORY " in the page of Device Adjust of the
Adjustments menu and by pressing ENTER.
duplicated.
2-1