Oxygen Sensors (Table 2) - Hach ORBISPHERE GA2 00 Series User Manual

O2 ec sensor
Hide thumbs Also See for ORBISPHERE GA2 00 Series:
Table of Contents

Advertisement

Table 5 Membrane specifications - Oxygen sensors (1) (continued)
Specification
Gaseous measurement range
Accuracy
Integrated radiation dose limit
[rads]
Expected current in air @ 1 bar
25 °C [μA]
Expected current in pure O
O
consumption in O
saturated
2
2
water at 25 °C [μg/hour]
Temp. compensation range
Temp. measuring range
5
Response time
Recommended min. liquid flow
6
rate
[mL/min]
Recommended min. linear flow
6
rate
[cm/sec]
Recommended gaseous flow rate
[L/min]

1.3.2 Oxygen sensors (Table 2)

Table 6 Membrane specifications - Oxygen sensors (2)
Specification
Recommended applications
Material
Thickness [μm]
Calibration gas
Dissolved measurement range
Gaseous measurement range
Accuracy
3
Accuracy is ± 0.1 ppb for 410, 510, 362x, 360x and 3655 instruments
4
Accuracy is ± 1 ppb for 366x and 3650 instruments
5
Response time at 25 °C for a 90% signal change
6
Liquid flow through an ORBISPHERE 32001 flow chamber, with protection cap and no grille
7
Halar is a trademark of Solvay Corporation.
2956A-A
0 Pa to 50 kPa
The greater of
±1% of reading
3
or ± 0.1 ppb
,
4
or ± 1 ppb
,
or ± 0.25 Pa
4
2 x 10
26.4
[μA]
132
2
40
7.2 sec.
180
200
2935A-A
Saturated to super
saturated levels
Halar
25
Air / Pure O
0 ppb to 400 ppm
0 Pa to 1000 kPa
The greater of
±1% of reading
or ± 10 ppb,
or ± 20 Pa
2958A-A
0 Pa to 100 kPa
0 Pa to 200 kPa
The greater of
The greater of
±1% of reading
±1% of reading
or ± 1 ppb,
or ± 2 Pa
8
10
9.4
47
14
– 5 to 60 °C
– 5 to 100 °C
9.5 sec.
120
100
0.1 to 3
29521A-A
Saturated to super
saturated levels
®
7
Tefzel
125
Air / Pure O
2
0 ppb to 400 ppm
0 Pa to 1000 kPa
The greater of
±1% of reading
or ± 10 ppb,
or ± 20 Pa
29552A-A
2952A-A
0 Pa to 200 kPa
The greater of
±1% of reading
or ± 2 ppb,
or ± 2 ppb,
or ± 5 Pa
or ± 5 Pa
N/A
10
6.3
5.4
31.4
27
9.4
90 sec.
38 sec.
50
50
30
30
2995A-A
In line hot wort
(up to 70 °C)
®
Tefzel
12.5
Pure O
2
0 ppb to 2000 ppm
0 Pa to 5000 kPa
The greater of
±1% of reading
or ± 50 ppb,
or ± 100 Pa
English 5
8
8
®
2