Proceed as follows to verify the Resistance Measurement function:
1. Connect the UUT to the 5500A as shown in Figure 4. Use a four-wire connection at
2. Set the UUT to the Resistance Measurement function.
3. Press [RANGE] on the UUT to lock on the 11Ω range.
4. Set the 5500A to the first test point in Table 6.
5. Observe that the reading on the UUT is within the range shown under the appropriate
6. Continue through the test points, being careful to lock the UUT on the specified
7. When you are finished, set the 5500A to [STANDBY].
NORMAL
V,
RTD
1000V
RMS
MAX
20V PK
MAX
TC
FLUKE 5500A CALIBRATOR
2-WIRE COMP ON
UUT
Range
11 ohm
11 ohm
110 ohm
110 ohm
1100 ohm
1100 ohm
11 kohm
11 kohm
16
the 5500A, transitioning to two wires at the UUT, and turn Two-Wire Compensation
on.
column.
range.
AUX
SCOPE
,
A,
-SENSE,
200V PK
AUX V
MAX
HI
20V
RMS
MAX
TRIG
LO
OUT
1V PK
MAX
20V PK
MAX
Figure 4. Resistance Measurement Verification Connections
Table 6. Resistance Measurement Verification Points
Minimum
Input
0.0 ohm
10 ohm
0.0 ohm
100 ohm
0.0 ohm
1000 ohm
0.0 ohm
10 kohm
V
RTD
30V
MAX
Maximum
1 Year
1 Year
-0.050
0.050
9.945
10.055
-0.05
0.05
99.90
100.10
-0.5
0.5
999.0
1001.0
-0.010
0.010
9.980
10.020
mA
mA
V
RTD
CAT
SOURCE
MEAS
300V
TC
MAX
30V
30V
MAX
MAX
FLUKE 74X
Minimum
Maximum
2 Year
-0.050
9.943
-0.05
99.88
-0.5
998.8
-0.010
9.980
gb05f.eps
2 Year
0.050
10.058
0.05
100.13
0.5
1001.3
0.010
10.020

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents