DC Volts Measurement

Proceed as follows to verify the DC Volts Measurement function:
1. Connect the UUT to the 5500A as shown in Figure 2. Connect the three lows (black
2. Set the UUT to the DC Volts Measurement function.
3. Press [RANGE] on the UUT to lock on the 110 mV range.
4. Set the 5500A to the first test point in Table 3.
5. Observe the reading on the UUT and check to see if it is within the range shown
6. Continue through the test points, being careful to lock the UUT on the specified
7. When you are finished, set the 5500A to [STANDBY].
NORMAL
V,
,
RTD
1000V
RMS
MAX
20V PK
MAX
_
TC
FLUKE 5500A CALIBRATOR
12
Ranges in the specification tables include the 10% over-range capability. Range
names on the 74X display do not include the 10% over-range. For example, the UUT
display shows
Range 100 mV,
Allow each item of verification equipment to satisfy its specified warm-up period.
Allow at least 5 minutes for warm up.
For each test, make sure the verification equipment has settled and that the
"unsettled" annunciator on the UUT is not displayed.
input jacks) with jumpers.
Do not force a dual banana plug between any two jacks in the
horizontal orientation. Doing so will damage the jacks. Use
the supplied jumper wires (P/N 944632). You can use a dual
banana plug in the vertical orientation.
under the appropriate column.
range.
AUX
SCOPE
A,
-SENSE,
200V PK
AUX V
MAX
HI
20V
RMS
MAX
TRIG
LO
OUT
1V PK
MAX
+
20V PK
MAX
Figure 2. DC Volts and AC Volts Measurement Connections
but the range name in the table is 110 mV.