Zero point can be adjusted manually in range -1.000 to +1.000 m
is keyed in, zero point adjustment will not be implemented.
See Reset mode - MAG 6000 SV (Page 90).
7.5

Service mode

All outputs of the transmitter can be forced-controlled in the service mode menu, see Service
mode (Page 91).
It is possible to check whether the outputs are functioning.
Error pending and status log lists are also accessible from this menu and the operating time (in
days) can be read.
The forced control is stopped and all previous settings are reinitialized the moment the service
mode is left by pressing top up key
7.6

MAG 5000 CT and MAG 6000 CT settings

Internal totallizers
Depending on the type of approval it is possible to reset the internal totalizers. The type of
approval is selected in the reset menu with the hardware key mounted. It is possible to choose
between:
● Hot/cold water
● Other liquids
Resetting of totalizers by electrical input is not possible.
Hot/cold water
● Totalizer 1 is allocated to forward flow (cannot be reset)
● Totalizer 2 is allocated to reverse flow (cannot be reset)
Other liquids
Both totalizer 1 and totalizer 2 are allocated to measure the net flow, i.e. any reverse flow will
make the totalizers count backwards.
● Totalizer 1 cannot be reset.
● Totalizer 2 can be reset if the flow velocity in the meter pipe is <0.25 m/s. When the totalizer
is reset, the pulse output register will also be reset.
Output
● When choosing hot water, changing the output settings is not allowed and the output setting
menus are not shown in display.
● When choosing cold water or other liquids, all output settings can be changed.
SITRANS FM MAG 5000/6000 IP67
Operating Instructions, 12/2019, A5E02338368-AD
7.6 MAG 5000 CT and MAG 6000 CT settings
.
Operating