Connecting

5.1 Electrical connection

5.1
Electrical connection
Power supply
Transmitter
115 to 230 V AC
Outputs
Passive output
(External powered)
Active output
(Powered from transmitter)
Relay output
Digital input
91 92 93 94 95 96 97
Sensor connection
1)
Shield
1)
Note:
Special cable with individual wire shields (shown as dotted lines) are only required
when using empty pipe function or long cables.
Figure 5-1
Note
Terminals 81 and 84 are only to be connected if special electrode cable with double screening
is used, e.g. when empty pipe function or long cables are used.
38
11 to 30 V DC/11 to 24 V AC
L1
1
L
N
N
2
PE
Current output
(Powered from transmitter)
31
32
Digital output
Vx 3 to 30 V
56
max. 110 mA
57
R
PLC-Digital input
58
24 V max. 30 mA
Counter or PLC-Digital input
56
57
R
58
44
NO
Relay
24 V DC/1A
45
NC
42 V AC/2A
46
Common
77
11 to 30 V DC
Input
78
Reserved for communication modules
Electrode cable
81
Brown
82
Black
0
Blue
83
84
Coil cable
Blue
85
Brown
86
Wiring diagram
PE
0/4 - 20 mA
Load ≤ 800 Ω
Menu setup
Negative:
Positive:
R =
Pull up/down resistor
1 to 10 KΩ may be
required - depending on
Cables/Input resistance
Sensor
81
82
0
83
84
85
86