Care list

What
Cleanable components
Carina basic device
Trolley
Compressed gas hose
Hinged arm
Reusable ventilation hoses
Reusable water trap

Dismantling the components

Switch off the device and respiratory gas
humidifier.
Disconnect mains plug.
1 Drain condensate from water traps.
2 Drain the water container of the humidifier.
Remove expiratory valve or leak valve. Remove
pilot line of the expiratory valve.
Remove the inlet filter of the basic device.
Respiratory gas humidifier, medication
nebulizer
Dismantle according to the relevant instructions
for use.
Ventilation hoses
Remove from the nozzles.
To remove the ventilation hoses, always
hold them at the connection sleeves and not
the spiral ribbing!
Otherwise the spiral ribbing may become
detached, e.g. at the connection sleeve, or the
hose can be torn out of the sleeve.
Drain condensate from ventilation hoses.
Disinfect and clean ventilation hoses, water
traps and their removed collecting jars.
Instructions for Use Carina SW 3.n
How often
Recommended cleaning
intervals
per patient
per patient
per patient/weekly
Cleaning, Disinfection and Sterilisation
How
Dismantle, clean and if applicable wipe
disinfect
Dismantle, clean and if applicable wipe
disinfect
Dismantle, clean and disinfect in the
cleaning and disinfection machine at
93 °C/10 minutes with detergents only.
Hot steam sterilisation at 134 °C.
2
1
105