Hose system with leak valve

1 Open the sliding cover on the underside of
Carina.
2 Using a coin, turn the switch to leak valve
»
« until it is properly engaged.
3 Close the sliding cover.
CAUTION
Switching between a hose system with leak valve
and a hose system with expiratory valve must be
carried out by properly trained personnel only!
Risk of malfunction!
4 Connect the filter to socket »
5 Connect the ventilation hose to the filter.
6 Connect the leak valve to the nozzle on the ven-
tilation hose, making sure that the direction is
correct.
When using the disposable hose system
VENTSTAR Carina LeakV (MP 00 312) the
leak valve is included.
Check that the hose system is displayed in the
status line of the screen as follows: »LeakV«.

Hose system with expiratory valve

1 Open the sliding cover on the underside of
Carina.
2 Using a coin, turn the switch to expiratory valve
»
« until it is properly engaged.
3 Close the sliding cover.
CAUTION
Switching between a hose system with leak valve
and a hose system with expiratory valve must be
carried out by properly trained personnel only!
Risk of malfunction!
Instructions for Use Carina SW 3.n
«.
2
4
5
2
Preparation
1
3
6
1
3
41