Installation

4.3.2 Installing the Reference Electrode

Figure 6
Installing the Teflon Beads
1
1.
Reference Electrode (Cat. No. 368429,00000)
2.
Reference Electrode Connector (Cat. No. 09186=A=0500)
*Teflon is a registered trademark of Dupont Corporation.
12
1. Place the seven small white Teflon
measuring cell as shown in
self-cleaning.
2. Remove the reference electrode from its packaging.
3. Remove the red cap from the electrode.
4. Connect the male end of the electrode to the female end of the red connector
on the measuring cell.
5. Pull the electrode from the electrode storage bottle.
6. Screw the electrode into the hole at the top of the measuring cell.
Note: Replace the reference electrode yearly. See
2
3
4
3.
4.
®*
beads through the hole at the top of the
Figure
6. The beads provide better mixing and
Section 7 on page
Teflon Beads (Cat. No. 588801,75008)
Measuring Cell (Cat. No. 09186=A=0100)
37.