Automatic Touch Hold Mode
To avoid electric shock, do not use the Touch Hold
mode to determine if a circuit with high voltage is
dead. The Touch Hold mode will not capture
unstable or noisy readings.
The Touch Hold mode automatically captures and displays stable
readings. Press R for 2 seconds. HOLD is displayed when
the meter is in the Touch Hold mode.
HOLD
2 seconds
+
When the meter captures a new input, it beeps and a new
reading is displayed.
Stray voltages can produce a new reading.
To exit the Touch Hold
measurement function. The Touch Hold mode and manual range
mode are mutually exclusive.

Bar Graph

The bar graph shows readings relative to the full scale value of a
displayed measurement range and indicates polarity.