5 Operation
5.4 Calibration of volume information
5.4
Calibration of volume information
Below is a step by step procedure for the calibration of the volume information transmit-
ted from the Peristaltic Pump P-1 instrument to the fraction collector when connected
to the remote control.
Step
1
2
3
4
Note:
Note:
48
Action
Set up the chromatography system as desired.
Connect Peristaltic Pump P-1 into the fraction collector with a communication
cable.
Set a flow rate approximately equal to that desired (see
flow rate, on page
45).
For further information refer to the instruction manuals for preferred fraction
collector, e.g. Frac 920.
The calibration remains valid whenever the Peristaltic Pump P-1 instrument
and the same fraction collector are connected via the remote outlet. If a differ-
ent i.d. tubing is used, the volume information should be recalibrated.
The calibration of the volume information is not valid after long term use of
the tube. Calibrate frequently to get high precision.
Peristaltic Pump P-1 Operating Instructions 28962201 AF
Section 5.1 Set the