Overview - Invacare Pronto M41 Series Service Manual

Hide thumbs Also See for Pronto M41 Series:
Table of Contents

Advertisement

Invacare® Pronto™ M41 Series
Component
Power legrests (if
fitted)
Battery mounting
Batteries
Remote / power
module
Chair configuration
Screws (every 6
months)

6.6 Overview

Under Rear Shroud/Seat
Depending on seat variant
(1) Power module
(2) Actuator module (optional)
24
Check
Remedy
Check cable
Replace cable if
necessary.
Check contacts
Check functions
Check battery support
Replace if necessary.
and mounting belts
for damage
Check batteries for
Replace batteries if
damage
necessary.
Check battery voltage
Charge batteries.
Check contacts and
Clean contacts and
terminals
terminals.
Remote, status LED
Evaluate error/flash
flashing
code.
Fixings
Tighten fixings,
replace if necessary.
Cables and connecting
Replace damaged
plugs
cables and tighten
connecting plugs, if
necessary.
Joystick function
Replace joystick knob
if necessary.
Replace remote if
necessary.
Power supply
Replace damaged
cables and tighten
connecting plugs, if
necessary.
Check chair
Update software
configuration version
if newer version
available.
Check screws for tight
Tighten screws if
fit
necessary.
Notes
See 6.12.5 Replacing Batteries,
page 43 and 6.12.2 Handling
Damaged Batteries Correctly, page
40.
See user manual.
See 6.12 Batteries, page 40.
See 6.4 Troubleshooting, page 20.
See LiNX service manual.
Fig. 6-1
1568396-C

Advertisement

Table of Contents
loading

Table of Contents