Agilent Technologies 16043A Operation And Service Manual page 59

3-terminal smd test fixture
Table of Contents

Advertisement

Figure 5-7
Attaching the Sliding Unit
Step 5. Secure the sliding mechanism with the pin (Agilent P/N 16043-24004) as shown in Figure
5-8.
Figure 5-8
Determining Sliding Mechanism Position
Step 6. Check that the sliding mechanism moves smoothly, and also that it can be fixed at
designated positions. If the mechanism moves abnormally, adjust the position of the ball
plunger while referring to "Adjusting Ball Plunger (16043A only)."
Chapter 5
Maintenance and Service
Replacement of Parts
59

Advertisement

Table of Contents
loading

This manual is also suitable for:

16043b

Table of Contents