Specifications and Supplemental Performance Characteristics
Supplemental Performance Characteristics
Figure 4-1
Impedance Additional Error
For D Measurement
The loss factor additional error De can be found using the impedance additional error Ze
[%] as shown in the following.
When Dx
When 0.1 < Dx
Dx is the measured value for D. However, this is not required when Ze is less than 10%.
NOTE
The additional error of D is not a percentage but an absolute value.
For Rs (ESR) Measurement
The additional error Rse [%] for Rs measurement can be found using the impedance
additional error Ze [%] as shown in the following.
When Dx
When 0.1 < Dx
Dx is the measured value for D. It is calculated using the following expression.
42
0.1
De = Ze / 100
0.5
De = ( Ze / 100)
0.1
Rse [%] = Ze / Dx
0.5
Rse [%] = ( Ze / Dx)
× π ×
×
Dx = 2
f
Here we have the following denotations:
×
(1 + Dx)
×
(
)

+

& Z
×
Csx
Rsx