To avoid electric shock, remove the power cord and any test
leads from the Meter before opening the current-input fuse
access door.
To replace the current-input fuses:
1. Unplug the power cord.
2. Turn the Meter upside down.
3. Remove the retaining screw on the fuse access door. See Figure 2.
4. Remove the protective cover from the fuse holders by slightly depressing the
back edge of the cover to unlatch it from the printed circuit board. Pull up on the
back edge of the cover and remove it from the fuse compartment.
5. Remove the defective fuse and replace it with a fuse of an appropriate rating. See
Table 3.
6. Replace the protective cover by pushing it over the fuses while aligning the
catches with the holes in the printed circuit board. Press the cover down until the
catches engage the printed circuit board.
7. Replace the fuse access door and install the retaining screw.
Fuses
Figure 2. Replacing the Current-Input Fuses
XW Warning
F2
F1
Digital Multimeter
Power Considerations
Bottom front
left corner
caw20.eps
7