4. Plug the small connector on the adapter cable into the side of the X-Rite DTP32 and
Small connector
AC adapter
5. Turn on the computer.
6. Calibrate the DTP32 (see the following section).
7. Calibrate the GA-1121 using ColorWise Pro Tools and the DTP32 (see
T
O CALIBRATE THE
1. Connect the densitometer to the computer and supply power (see
2. From the Main Menu on the X-Rite DTP32 display, press the p1 key once to reach p2.
3. Press the cal key.
4. Insert the X-Rite Auto-Cal Strip into the 35mm slot on the front of the X-Rite DTP32—
plug the AC adapter into a wall outlet.

Calibrating the DTP32

You will need the black-and-white X-Rite Auto-Cal Strip included with the
densitometer. Calibrating the densitometer does not require ColorWise Pro Tools.
X-R
DTP32
ITE
is displayed, followed by the words INSERT CAL STRIP .
Calibrating motor speed
arrow end first—until it stops or the densitometer starts pulling the strip.
Reading
appears momentarily, followed by the density values and CALIBRATION OK .
The densitometer automatically returns to the MAIN MENU .
If UNRECOGNIZABLE STRIP appears, repeat the process or try cleaning the strip (see the
X-Rite DTP32 Operating Manual).