Installing / removing and connecting analyzer and option modules
7.8 Wiring option modules
Procedure
1. Connect the option modules as follows after installation:
– OM 1.1 to the processing unit using the 50-pin ribbon cable
– OM 2.x: Use the following cables to connect the OM2.x:
2. Connect the option modules with the corresponding terminal block:
76
Option
Cables
module
OM 2.1
26-pin ribbon cable with the analyzer module PCB(s)
OM 2.2
26-pin ribbon cable with the analyzer module PCB(s) and via the 12-pin ribbon
cable (CAN BUS) with the processing module
Connection of the 26-pin ribbon cable to the AM1
Connection of the 26-pin ribbon cable to the AM2
Figure 7-14
Position of the connector sockets based on example of OM2.1
Operating Instructions, 05/2018, A5E31930403-06
Wall-mounted device