Combined operation
5.4 Application planning
5.4.6

Operation with different analyzer modules

OXYMAT 7/ ULTRAMAT 7/ CALOMAT 7
Table 5- 4
Operating
mode
Pressure
operation
Suction op-
eration
44
Mounting conditions and positions for the sample gas restrictor
Sample gas restrictor/clamping screw
OXYMAT 7/ ULTRAMAT 7 ULTRAMAT 7/ CALOMAT 7 OXYMAT 7/ CALOMAT 7
OXYMAT 7 must always
be flowed first.
Install a sample gas re-
strictor in front of the sam-
ple gas inlet of the
OXYMAT 7.
ULTRAMAT 7 must always
be flowed first.
Install a sample gas re-
strictor in the sample gas
outlet of the OXYMAT 7.
Replace a sample gas
restrictor in the sample gas
inlet of the OXYMAT 7 with
a clamping screw.
We recommend flowing the
CALOMAT 7 as the first
module
Operating Instructions, 05/2018, A5E31930403-06
No restrictions
Wall-mounted device