3. Loosen the three screws
4. Loosen the screw
bottom of the enclosure.
5. Remove the complete cable gland set.
6. Install the new complete cable gland set
Note
Fastening the cable glands
Fasten the cable glands
• M16 x 1.5 clamping range 5 to 10 mm made of plastic, counter nut and cap nut (clamp
• M20 x 1.5 clamping range 10 to 14 mm made of plastic, counternut and cap nut with a
• M16 x 1.5 clamping range 6 to 10 mm made of metal, counternut and cap nut with a
• M20 x 1.5 clamping range 10 to 14 mm made of metal, counternut and cap nut with a
7. Tighten the screw
8. Remount the shield plate. The torque at the screws
9. Close the door using six screws.
Wall-mounted device
Operating Instructions, 05/2018, A5E31930403-06
on the inside of the right side panel and the eleven hex nuts at the
cables or plugs) with a torque of 3.0 Nm.
torque of 4.5 Nm.
torque of 6.5 Nm.
torque of 9.0 Nm.
again with a torque of 2.0 Nm.
and remove the shield plate
with eleven hex nuts with a torque of 0.8 Nm.
Maintenance and servicing
11.9 Replacing the cable gland set
.
amounts to 0.8 Nm.