RUSKA 2482
Users Manual
Specifications
Differential Pressure Range
•
Line Pressure Range
•
Accuracy
•
Accuracy is defined as the expended uncertainty in pressure determined using the method
described in ISO "Guide to the Expression of Uncertainty in Measurement," and
represents an approximate 95% level of confidence. The accuracy capability of 40 ppm of
reading plus 0.009 kPa, combined root-sum-of-squares, is applicable with corrections for
actual mass values and when reference conditions.
Differential Pressure Reproducibility
•
Refer to the Calibration Report for Pressure Uncertainty.
Piston and Cylinder
•
•
Electrical Power Requirements
•
•
Operating Conditions
•
•
Storage Conditions
•
•
Instrument Conditions
•
Instrument Weight
•
Operating Medium
•
•
36
0 to 29 psi (0 to 200 kPa)
0 to 2900 psi (0 to 200 bar)
40 ppm of reading +0.009 kPa
Differential Pressure Reproducibility: +/-4.5 Pa
Thermal Coefficient (piston + cylinder): 9.2 x 10
Piston & cylinder material: tungsten carbide
12v DC 5A max
Universal power supply included — 120/240 VAC 50/60 Hz
64 °F to 92 °F (18 °C to 28 °C)
20% RH to 75% RH, non-condensing
32 °F to 122 °F (0 °C to 50 °C)
0% RH to 90% RH, non-condensing
18 in x 15 in x 18 in (48 cm x 38 cm x 48 cm)
Approximately 88 lb (40 kg)
Internal operation medium: Dioctyl sebacate (DOS)
Test item operating fluid: Dioctyl sebacate (DOS) or clean dry gas
1
-6
-1
˚C