34134A AC/DC DMM Current Probe The 34134A Current Probe measures low DC or AC from 10 mA to 50 A DC, 40 A AC. This battery-powered clamp-on probe may be used with digital multimeters, voltmeters, or other voltage-measuring instruments. The probe jaws clamp around the conductor under test, allowing current measurements without breaking the circuit.
34134A AC/DC DMM Current Probe Specifications and characteristics Specifications and characteristics Operating and environmental characteristics of the 34134A Current Probes are shown below. Electrical Specifications 1 V/A Range (1 mV/mA) 10 mV/A Range Current Range DC: 10 mA to 2 A DC: 100 mA to 80 A AC: 10 mA to 1.5 A AC: 100 mA to 60 A...
34134A AC/DC DMM Current Probe Specifications and characteristics Typical Frequency Response Curve Voltage vs Frequency Response Curve Environmental Characteristics Operating Temperature 0 to 50°C Storage Temperature -30 to 80°C Operating relative 0 to 50 °C, 0 to 85% relative humidity humidity Temperature Coefficient 800 ppm per °C +10 mA/°C typical from -10 to 50°C Altitude...
34134A AC/DC DMM Current Probe Specifications and characteristics Safety Specifications Working voltage 600 V CAT III Case protection IP20 per IEC 529 Drop test 1m according to IEC 68-2-32 Mechanical shock 100 G: test per IEC 68-2-27 Vibration Test per IEC 68-2-6 frequency range 10 Hz to 55 Hz, amplitude: 0.15mm This symbol signifies that the 34134A AC/DC DMM Current Probe is protected by double or reinforced insulation.
34134A AC/DC DMM Current Probe Operation Operation 1 Connect the probe to the multimeter or other instrument. CAUTION Be sure to observe the polarities: red = positive terminal, black = common terminal. 2 Select the multimeter range which best corresponds to the measured current. For better reading stability, you may use the DMM 2 Volt range and let the probe “warm up”...
34134A AC/DC DMM Current Probe Calibration Calibration To guarantee that your instrument complies with the factory specifications, we recommend that the 34134A Current Probe be submitted to your local Agilent Service Center at one year intervals for recalibration, or as required by other standards. Repair This probe contains no user serviceable parts.