DC Current (Rear Input)

Configuration: 7.5 Digits, Internal Guard, Rear Input, Filter Off, Aperture = Auto
Range
Reference Value
100 μA
0.000 00 μA
100 μA
100.000 00 μA
100 mA
0.000 00 mA
100 mA
100.000 00 mA

AC Voltage (Rear Input)

Configuration: Input: Rear, Ext Guard: Off, Resolution: 6 digits, Measure Setup: DC 1 MΩ, RMS Filter
40 Hz, Frequency Path Bandwidth Limit: OFF, Band: Wideband
Suggested Reference: 5730A
Range
Frequency Reference Value
1 V
1 kHz
10 V
1 kHz

AC Current (Rear Input)

Configuration: 7.5 Digits, Internal Guard, Front Input, DC Coupled, Wideband RMS Filter 40 Hz.
Suggested Reference: 5730A
Range
Frequency Reference Value
100 μA
1 kHz
100 μA
10 kHz
Resistance – HV Ohms (Rear Input)
Configuration: 7.5 Digits, Aperture = Auto, 4-wire, Internal Guard, Rear Input, Filter Off.
Suggested Reference: 10 GΩ Standard Resistor
Range
Reference value
10 GΩ
0.000 000 GΩ
10 GΩ
10.000 000 GΩ
Permissible Error (±)
8588A
0.000 50 μA
0.001 60 μA
0.001 30 mA
0.005 60 mA
1.000 000 V
0.000 089 V
1.000 00 V
0.000 89 V
100.000 0 μA
0.040 0 μA
100.000 0 μA
0.072 3 μA
Permissible Error (±)
8588A
0.012 900 GΩ
0.019 677 GΩ
Reference Multimeter / 8½ Digit Multimeter
8558A
0.000 60 μA
0.001 90 μA
0.000 60 mA
0.006 80 mA
Permissible Error (±)
8588A
8558A
0.000 100 V
0.001 00 V
Permissible Error (±)
8588A
8558A
0.052 9 μA
0.091 6 μA
8558A
0.012 900 GΩ
0.019 677 GΩ
Calibration Verification
Fluke Calibration
Uncertainty (±)
0.000 12 μA
0.000 45 μA
0.000 26 mA
0.000 58 mA
Fluke
Calibration
Uncertainty (±)
0.000 018 V
0.000 17 V
Fluke
Calibration
Uncertainty (±)
0.006 6 μA
0.008 1 μA
Fluke
Calibration
Uncertainty (±)
0.000 010 GΩ
0.001 600 GΩ