Relays in
Optional Relay
Analog
Digital
Communications Modbus / TCP Ethernet / Power over Ethernet (PoE)
Certification and Specification
Performance
Transport System
Flow Rate
Transport Time
Sample Line
Tubing
Tubing Length
Exhaust Line
Tubing
Exhaust Length
Ambient Point
Operating Temperature
Wiring Requirement
4-20mA
Digital
Gas Concentration Display and Interface
Instrument
Midas-M
Specifications
"8 relays for Alarm1 and Alarm2 of each channel and a relay for
common fault rated 1.0 A @ 30Vdc or 0.5 A @ 125 Vac Max 10uA @
10mV minimum, configurable as normally open or closed, latched or
unlatched"
3 wire sink, 3 wire source, or 4 wire fully isolated; 0 to 21 mA for each
channel
"CE marked Meets EN 50270:2015+AC:2016 and EN61000-6-
4:2007+A1 ETL approved UL 61010-1:2012 Ed.3 IEEE 802.3af-2003"
Refer to Individual sensor datasheets
600 mL/min
2 to 25 seconds maximum
3.18 mm ID X 6.35 mm OD (0.125 X 0.25 in)
Up to 30 m (100 ft) with FEP tubing
4.76 mm ID X 6.35 mm OD (0.188 X 0.25 in)
Up to 30 m (100 ft)
In line air filter required
0°C to +40°C (32°F to 104°F)
2 wire, 14 AWG maximum
CAT5 Cable or equivalent: RJ45 connector
"4-digit alphanumeric display with separate units, concentration bar
graph for each channel and other icon driven indicators 4 button
interface keypad"
75