Installation

CAUTION: READ SECTION 5 ENTITLED BATH USE before placing the
bath in service. Incorrect handling can damage the bath and void the war-
ranty.
This section provides the steps required to set up and operate the 6055 bath.
This section should be used as a general overview and reference and not as a
substitute for the remainder of the manual. Please read Sections5 through7
carefully before operating the bath.
WARNING: Never move a bath that is full of "hot" fluid. This action is
extremely dangerous and can result in personal injury and possible dam-
age to the surrounding area.
The bath comes on a cart. This cart is for ease of operation, but in no way im-
plies that the bath is portable. This bath is extremely dangerous when filled
with salt. Follow the safety guidelines set forth in the Before you start section.
The bath should never be picked up with liquid salt in the tank. If the bath
needs to be picked up, allow the salt to solidify and cool. Four people are
needed to safely lift the bath. Lift the bath where the bath and the cart meet.
CAUTION: Use caution as this bath is top heavy.
4.1

Unpacking

Unpack the bath carefully and inspect it for any damage that may have oc-
curred during shipment. If there is shipping damage, notify the carrier
immediately.
Verify that the following components are present:
6055 Bath
Control Probe
Power Cord
Immersion Heater
Report of Test