Before using any cleaning or decontamination method except those rec-
ommended by Hart, users should check with an Authorized Service Cen-
ter to be sure that the proposed method will not damage the equipment.
If the instrument is used in a manner not in accordance with the equip-
ment design, the operation of the bath may be impaired or safety hazards
may arise.
The over-temperature cut-out should be checked every 6 months to see
that it is working properly. In order to check the user selected cut-out, fol-
low the controller directions (Section 8.2) for setting the cut-out. Both the
manual and the auto reset option of the cut-out should be checked. Set the
bath temperature higher than the cut-out. Check to see if the display
flashes cut-out and the temperature is decreasing.
WARNING: When checking the over-temperature cut-out, be sure that the
temperature limits of the bath fluid are not exceeded. Exceeding the tem-
perature limits of the bath fluid could cause harm to the operator, lab, and
instrument.
The constant temperature bath depends upon the certain qualities of the
fluid medium in order to maintain a uniform and stable temperature envi-
ronment.