Figure 4-13. NI 6517 Output Signal Connection Example
+3.3 V
Photocoupler
Caution
power supply and that GND is connected to the negative pole of the power supply.
Failure to do so could permanently damage the NI 6519 device and the power supply
if it is not protected.
Figure 4-14. NI 6519 Output Signal Connection Example
+3.3 V
Photocoupler

Distributing Current (NI 6517/6519 Only)

In applications that drive many high-current loads, a large amount of current must be returned
on the GND pins. When connecting the NI 6517/6519 to multiple loads, physically connect to
all GND pins on the device, as shown in Figure 4-15 and Figure 4-16. Connecting to all GND
pins distributes the current evenly among the GND pins and lowers the amount of current driven
on any single cable wire and on the accessory terminals.
Voltage
Regulator
Output
Darlington
Driver
NI 6517
Make sure that OUT.COM (VCC) is connected to the positive pole of the
Voltage
Regulator
Output
Darlington
Driver
NI 6519
COM (VCC)
P<0..3>.<0..7>
+5 V ~ +30 V
GND
OUT.COM (VCC)
P<2..3>.<0..7>
+5 V ~ +30 V
GND
© National Instruments | 4-9
NI 651x User Manual