Detailed Specifications

Normal-Sinus-Rhythm Waveform

ECG Reference ...................................................... The ECG amplitudes specified are for Lead II (calibration), from the baseline to the peak
Normal Sinus Rhythm ........................................... 12-lead configuration with independent outputs referenced to right leg (RL). Output to 10
Amplitude ............................................................... 1.0 mV. Other leads are proportional to Lead II (reference lead) in percentage per:
Lead I .................................................................. 70
Lead II ................................................................. 100
Lead III ................................................................ 30
Lead V1 ............................................................... 24
Lead V2 ............................................................... 48
Lead V3 ............................................................... 100
Lead V4 ............................................................... 120
Lead V5 ............................................................... 112
Lead V6 ............................................................... 80
Amplitude Accuracy .............................................. ±5 % of setting Lead II
ECG Rate ................................................................ 30, 60, 80, 90, 120, 150, 180, 210, 240, 270, 300, and 320 BPM (Preset Hypotensive
Rate Accuracy ....................................................... ±1 % of setting
ECG Waveform Selection ..................................... Adult (80 ms) or neonatal (40 ms) QRS duration
Power-On Default .................................................. 60 BPM, 1.0 mV, adult QRS

Arrhythmia

Atrial Fibrillation .................................................... Coarse or fine
Premature Ventricular Contraction ..................... Left Ventricular
Ventricular Tachycardia ........................................ 160 or 200 BPM
of the R wave. All other leads are proportional.
Universal ECG Jacks, color-coded to AHA and IEC Standards.
condition is at 40 BPM)