Invasive Blood Pressure; Non-Invasive Blood Pressure - Fluke ProSim 4 Getting Started

Vital signs simulator
Table of Contents

Advertisement

Invasive Blood Pressure

Channels ................................................................ 1 electrically isolated from all other signals
BP Output ............................................................... Circular DIN 5-Pin
Input/output Impedance ........................................ 300 Ω ±10 %
Exciter Input Range ............................................... 2.0 to 16.0 V peak
Exciter-Input Frequency Range ........................... DC to 5000 Hz
Transducer Sensitivity .......................................... 5 μV/V/mmHg
Pressure Accuracy ................................................ ±(1 % of setting + 1 mmHg) Accuracy guaranteed for DC excitation only
Static Pressure ...................................................... 0, 80, 160, and 250 mmHg
Dynamic Waveforms
Synchronization ................................................... To ECG heartrate
Chambers simulated systolic/diastolic pressure:
Type
Adult
Adult
Adult
Adult
Neonatal
Neonatal

Non-Invasive Blood Pressure

Pressure Units ....................................................... mmHg
Manometer (Pressure Meter)
Range .................................................................. 10 to 400 mmHg
Resolution ........................................................... 0.1 mmHg (for display purposes)
Accuracy.............................................................. ±(1 % reading +1 mmHg)
Pressure Source .................................................... Inflation bulb or device under test
IBP (Arterial)
IBP (Left Ventrical)
60/30
60/0
120/80
120/0
150/100
150/0
200/150
200/0
35/15
35/0
70/40
70/0
Vital Signs Simulator
Detailed Specifications
19