Commissioning
7.4 Fast commissioning (Quick Start) with SIWATOOL
● Select "Service mode off" in the Service menu.
⇒ The calibration is finished and the correct weight value is now displayed in the SIWATOOL
program.
Receiving calibration parameters
During the calibration, the SIWAREX internally changes its calibration digits. This means that
SIWATOOL now has obsolete parameter values. Obsolete calibration parameters are marked
in red.
The calibration parameter (DR3) must be read back in order once again obtain the calibration
digits in the SIWATOOL which are consistent with the SIWAREX.