Section 5
Figure 18
Attaching Portable AquaTrend Interface to Sensor
Page 58
Portable AquaTrend Interface Installation and Mounting
The Portable AquaTrend Interface can be easily connected to a sensor by
plugging the cord jack on the Portable AquaTrend Interface into the receptacle
provided on the sensor. See
The Portable AquaTrend Interface's mounting kit provides an easy way to
mount the device to a wall or pole. See
instructions which follow.
For wall mounting:
1. Use the bracket as a template to mark positions on the wall for four of the
eight screw holes – either the vertical or horizontal pair.
2. Drill the marked positions for four ¼-20 screws.
3.
Use four ¼-20 hex-head or pan-head screws to secure the bracket to the
wall (10 in-lb (1.1 n-m) maximum).
4. Fit the bracket lip into the slot on the back of the Portable AquaTrend
Interface and push down until the device fits securely in the bracket.
Figure
18.
Figure 19
and
51350-18 Hardware Installation.fm
Figure 20
and the