Hach AquaTrend 51200 Instrument Manual page 42

Table of Contents

Advertisement

Section 3
Page 42
Restore Net Connections
9. Select LIST NETWORK CONNECTIONS and press the ENTER key. All
connected devices will be shown in the resulting display. Scroll through
the list and ensure all devices were restored onto the network.
Note: If a device still is not shown on the display, add it using the Add Device
procedure given in
section 3.4.2 on page
10. Press the EXIT key to return to the previous menu. Select ALARM TO
RELAY and check/correct all displayed information.
11. Press the EXIT key to return to the previous menu. Select RECORDER
TO ANALOG OUT and check/correct all displayed information.
12. Press the EXIT key to return to the previous menu. Select SENSOR TO
DISPLAY and check/correct all displayed information.
After the Restore Net Connections procedure is completed and all
devices are present on the network, you must:
1. Reset the date and time (see
2. Reattach measurements to channels (see
3. Redefine graph settings (see
4. If SOM parameters such as comm loss, range, and trim (if range is set to
other than the default) were customized, those custom parameters will
need to be reestablished. Please refer to the SOM Manual ,
Cat. No. 51250-18.
26.
section 3.3 on page
24).
section 3.4.4 on page
)
section 3.5.2 on page 32
51350-18 Initial Aquatrend Setup.fm