Pcu Assembly - 10 Mm Nominal Diameter - Vacuum; Piston Disassembly; Piston Cleaning - Fluke P3000 Series User Manual

Pneumatic deadweight tester
Hide thumbs Also See for P3000 Series:
Table of Contents

Advertisement

Item
PCU Assembly — 10 mm Nominal Diameter — Vacuum

Piston Disassembly

Piston Cleaning

1.800.561.8187
Figure 5-1. PCU Assembly - 10mm - Pressure
Table 5-1. PCU Assembly - 10 mm - Pressure - Parts List
Description
1
Weight Carrier
2
Cylinder
1. Unscrew the PCU assembly from the instrument; use the dowel hole if the cylinder is
tight.
2. The weight carrier is permanently attached to the piston, hold the weight carrier and
gently withdraw the piston from the cylinder (2)
3. Use "non-fluffing", non-abrasive, lint-free tissue or absorbent cloth. Hold the piston
by the weight carrier, and rub the tissue back and forth along its length.
4. To remove all traces of contamination, the piston can be cleaned in a suitable solvent.
O-ring seals (where fitted) are Nitrile rubber, and should not be
immersed in solvents, as they will become damaged. They
should be wiped carefully with a new tissue.