Switching Output Voltage; Output Voltage Fine Adjustment - A&D GX Series Instruction Manual

Hide thumbs Also See for GX Series:
Table of Contents

Advertisement

Note
"Full scale" of the full scale output mode indicates the full scale values shown in the tables
above. The output voltage may exceed 1.000 V, depending on the weighing data.
For example, when the GX-2000 displays 2100 g, the output voltage is 1.05 V.

5-3 Switching Output Voltage

The output voltage can be switched using the slide switch on the GX-06 panel. "0V〜" has been set
at factory before shipment.
"0V〜" (0-1 V):
At zero 0.000 V
"0.2〜" (0.2-1 V):

5-4 Output Voltage Fine Adjustment

The output voltage has been adjusted at the factory before
shipment.
Using the ZERO and SPAN fine-adjustment controls and
a voltmeter, output voltage can be fine adjusted.
Fine-adjustment procedure
1. Turn the display off. At this time, the output voltage
will be at zero. Turn the ZERO control so that the
voltmeter indicates 0.000 V when the slide switch is
set to "0V〜"; 0.200 V when the slide switch is set
to "0.2V〜".
2. While pressing and holding the SAMPLE and
PRINT keys, press the ON:OFF key. The balance
displays p5
At full scale
At zero 0.200 V
. At this time, a voltage of 1 V is
1.000 V
At full scale
31
1.000 V