Trim The Transmitter; Zero Trim Via Class 2 Master - Emerson Rosemount 3051 Series Quick Start Manual

Pressure transmitter/flow meter with profibus pa protocol
Hide thumbs Also See for Rosemount 3051 Series:
Table of Contents

Advertisement

February 2019
6

Trim the transmitter

Devices are calibrated by the factory. Once installed, it is recommended to
perform a zero trim on the sensor to eliminate error due to mounting position
or static pressure effects.
This can be done by performing a zero trim via:
LOI – see
Class 2 master – see
6.1

Zero trim via Class 2 Master

Procedure
1. Place the transducer block into Out of Service (OOS) mode.
2. Apply zero pressure to device and allow to stabilize.
3. Go to Device Basic Setup > Calibration and set the Lower Calibration
Point to 0.0.
4. Place the transducer block to AUTO mode.
Quick Start Guide