11 Configuration

Valve Box

The valve box mounts on top of the column oven. It may contain up to six valves mounted on
heated blocks. Each block can accommodate two valves.
Valve positions on the blocks are numbered. We suggest that valves be installed in the blocks
in numeric order.
All heated valves in a valve box are controlled by the same temperature setpoint.
To Configure a Valve:
1 Select Settings > Configuration > Valves.
2 For each valve installed, select the valve type, then set the loop volume, step time, and BCD
3 Click Apply to store the changes.
186
inversion as applicable to the valve type.