National Instruments USB-6008 User Manual page 6

Bus-powered multifunction daq usb device
Hide thumbs Also See for USB-6008:
Table of Contents

Advertisement

Choose one of the labels, align the correct label with the terminals printed on the top
panel of your device and apply the label, as shown in the previous figure.
1
2
3
1. User-Defined Custom Label
2. Terminal Number Label
3. Digital I/O Label
Note
them into the matching connector jack, as indicated by the overlay label on the
device.
5.
Plug one end of the USB cable into the NI USB-6008/6009 and the other end into an
available USB port on the computer.
6.
Double-click the NI MAX icon on the desktop to open Measurement & Automation
Explorer (MAX).
7.
Expand My System»Devices and Interfaces and verify that the NI USB-6008/6009 is
listed. If your device does not appear, press <F5> to refresh the view in MAX. If your
device is still not recognized, refer to
information.
8.
Self-test your device in MAX by right-clicking NI USB-6008 or NI USB-6009 and
selecting Self-Test. Self-test performs a brief test to determine successful device
installation. When the self-test finishes, a message indicates successful verification or if
an error occurred. If an error occurs, refer to
Caution
with shielded cables and accessories.
9.
Connect the wires (16 AWG to 28 AWG) of a shielded, multiconductor cable to screw
terminals by stripping 6.35 mm (0.25 in) of insulation, inserting the wires into the screw
terminals, and securely tightening the screws with the flathead screwdriver to a torque of
0.22 N · m to 0.25 N · m (2.0 lb · in. to 2.2 lb · in.). Refer to the
Descriptions
section for an image of the NI USB-6008/6009 pinout.
If using a shielded cable, connect the cable shield to a nearby GND terminal.
Note
information about IEEE 1451.4 TEDS smart sensors, go to
10. Run a Test Panel in MAX by right-clicking NI USB-6008 or NI USB-6009 and selecting
Test Panels.
Figure 3. NI USB-6008/6009 Signal Labels
+
-
+
-
AI1
GND
AI0
GND
GND
GND AI0 AI4 GND AI1 AI5 GND AI 2 AI6 GND AI3 AI7 GND AO0 AO1 GND
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
ANALOG
32 31 30 29 28 27 26 25 24 23 22 21 20 19 18 17
GND +5V+2.5V PFI0 P1.3 P1.2 P1.1 P1.0 P0.7 P0.6 P0.5 P0.4 P0.3 P0.2 P0.1 P0.0
After you label the screw terminal connector plugs, you must only insert
To ensure the specified EMC performance, operate this product only
For information about sensors, go to
+
-
+
-
AI2
GND
AI3
GND
AO0 AO1 GND
DIGITAL
4. Analog Input Differential Signal Name Label
5. Analog Input Single-Ended Signal Name Label
ni.com/support/daqmx
ni.com/support/daqmx
ni.com/sensors
NI USB-6008/6009 User Guide | © National Instruments | 5
4
or
5
1
for troubleshooting
.
Pinout and Signal
. For
ni.com/teds
.

Advertisement

Table of Contents
loading

This manual is also suitable for:

Usb-6009

Table of Contents