Installation/Sequence Of Operation - Invacare 6LXO2 Service Manual

With senso2
Table of Contents

Advertisement

INSTALLATION / SEQUENCE OF OPERATION
MODEL 3LX and MODELS 5LX,5LXO2, 5LXO2T, 6LXO2, 6LXO2T
INSTALLATION / VERIFICATION OF BATTERY FREE
When your new Invacare concentrator arrives, it should be checked for proper operating conditions.
1. If the unit has been in below-freezing temperatures, allow it to warm up to room temperature (approxi-
mately 30 minutes) before operating.
2. The concentrator will need to be turned on for 4 to 5 seconds to charge the Battery Free
Connect power cord to outlet and turn the concentrator on. Turn flow control knob counterclockwise and flow will
begin immediately. Set flow rate of the Invacare 3 to 3 L/min., Invacare 5 to 5 L/min. and the Invacare 6 to 6 L/min.
Turn unit off.
3. Unplug the power cord and press On/Off switch to the ON position. An intermittent audible alarm will
sound. This confirms proper operation of the Battery Free
OFF.
4. Connect cord to outlet and turn on concentrator. Unit will sound 3 BEEPS on startup.
5. Check the oxygen concentration per specifications after 30-40 minutes running time.
SEQUENCE OF OPERATION
Applying 120 VAC via the power switch energizes the compressor motor, hour meter, cooling fan and the
printed circuit (P.C.) board.
Room air enters the compressor via the cabinet filter, compressor inlet filter, and sound muffler. The air is
compressed by the wobble pistons in the compressor to a pressure of 20 PSI (137.72 kPa) BEFORE
Serial No. "97K" or 21 PSI (144.79 kPa) AFTER Serial No. "97K".
As increased pressure creates increased temperature, a heat exchanger is utilized to lower the tempera-
ture by 20
O
F (6.7
O
C) before the air enters the 4-way valve. It is then channeled to a sieve bed containing
the adsorption material. A restrictor downstream of the sieve bed causes pressure to build up inside the
sieve bed which is necessary for the adsorption process. A small amount of relatively pure oxygen enters
the top of the second bed with the balance entering a storage tank. The nitrogen removed is exhausted
back from the bed through the four-way valve into room air. A muffler is located at the exhaust end of the
valve to muffle the sound of the exhaust as it exits the concentrator.
The oxygen not being used to exhaust is channeled into the storage tank. The pressurized oxygen is
regulated down to 5 PSI (34.4 kPa), enters an accurate flow-measuring device, flows through a bacterial
filter, a check valve and out to the patient.
The electrical activation of the 4-way valve is accomplished every 8 to 21 seconds by the pressure sensor
and P.C. board electronics when the pressure set point of 20 + 3 PSI (137.72 + 51.32 kPa) BEFORE
Serial No. "97K" or 21 + 3 PSI (144.79 + 51.32 kPa) AFTER Serial No. "97K" is reached. The time
between cycles is dependent on altitude, flow rate and internal environmental factors.
On the 5LX/5LXO2/5LXO2T/6LXO2/6LXO2T, a P.E. or Pressure Equalization valve opens just prior to
the shift of the 4-way valve. This allows highly concentrated oxygen to enter the just exhausted bed from
the top. This additional pressure allows the bed to start its cycle at a higher pressure. The P.E. valve will
close just after the shift of the 4-way valve.

INSTALLATION/SEQUENCE OF OPERATION

TM
POWER LOSS ALARM
TM
Power Loss Alarm. Turn On/Off switch
11
N
S
T
A
L
L
TM
Power Loss Alarm.
A
T
O
N
Q
U
N
C
O
O
R
O
N
I
I
S
E
E
E
F
P
E
A
T
I

Advertisement

Table of Contents
loading

This manual is also suitable for:

5lxo2t3lx5lx5lxo26lxo2h6lxo2t ... Show all

Table of Contents