Adjusting Fork Rotation

Caster angle vertical axis MUST be at 90° +/- 1° from the ground, if not a
flattering risk may occur. Both caster assemblies should be at the same angle.
After ANY adjustments, before use, make sure all attaching hardware
is tightened securely - otherwise injury or damage may occur.
NOTE: For this procedure, refer to FIGURE 5.8.
1. Remove the dust cover from the caster head tube.
2. To properly tighten caster journal system and guard
against flutter, perform the following check:
A. Tip back the wheelchair to floor.
B. Pivot caster fork assemblies in rear position
C. Push caster to initiate clockwise movement
D. Let caster fork assembly rotate freely,
if not change the upper specific bearing.
3. Adjust locknuts to appropriate torque (50/60 Nm)
with a 19 mm socket wrench, do not overtight
the locknut.
4. Test wheelchair for manoeuvrability.
5. Snap dust cover into the caster head tube.
6. Perform STEPS 1 to 5 on the other side.
SECTION 5—REAR WHEELS/FRONT CASTERS
W A R N I N G
Locknut
Specific
Bearing
Caster
Head tube
FIGURE 5.8
3 5
Dust
cover
Adjusting Fork Rotation
Fork