SECTION 5—OPERATING INSTRUCTIONS
GREEN light - with YELLOW light flashing - Call supplier
IMMEDIATELY. Oxygen sensor malfunctioning; you may continue
to use the concentrator (FIGURE 5.7).

Indicators (LX Model)

NOTE: For this procedure, refer to FIGURE 5.8.
Explanation of LX models Indicator Lights
RED light ( ) - Total Unit Shut-Down. Switch IMMEDIATELY to a
back-up oxygen supply. Call supplier IMMEDIATELY.
GREEN light (I/O) - On/Off. System okay.
Auto Shut Down
RED
Use Backup
Call Supplier
I/O
Normal
GREEN
FIGURE 5.8 Indicators (LX Model)

Main Power Loss Alarm System

The alarm system contains no battery - and is thereby maintenance
free. The alarm system is powered by a capacitor that is
continuously charged and ready to trigger the alarm in the event of
main power loss.

Elapsed Time Meter (Hour Meter)

The hour meter displays the cumulative number of hours the unit
has operated.
Part No 1141493