SECTION 21—PURITY INDICATORS
SECTION 21—PURITY INDICATORS
SensO
Alarm Indicators
2
DANGER
To prevent electrical shock, ALWAYS disconnect concentrator from electrical outlet 
before servicing.
NOTE: For this procedure, refer to FIGURE 21.1 on page 72.

Operation

The Perfecto
V with SensO
Concentrator is equipped with an oxygen purity indicator. 
2
2
 
This feature monitors the purity level of the oxygen generated by the concentrator. If 
purity falls below alarm thresholds, indicator lights on the control panel will illuminate. 
NOTE: Concentrator may be used during the initial start warm‐up time (approximately 30 
minutes) while waiting for the O
purity to reach maximum.
When the unit is turned on, the GREEN light will come on (SYSTEM OK/O
 greater than
2
 
85%). After five minutes, the oxygen sensor will be operating normally and will control 
the indicator lights depending on oxygen concentration values. The explanation of the 
indicator light functions are as follows:

Indicator Lights

NOTE: Indicator lights are visible only when lit.
O
GREEN light (
) ‐ Normal Operation.
2
YELLOW light (
)‐ Call supplier IMMEDIATELY. You may continue to use the 
concentrator unless instructed otherwise by your supplier. Be certain that backup oxygen 
is nearby.
RED light (
)‐ Total unit shutdown. Switch IMMEDIATELY to backup oxygen supply. 
Call supplier IMMEDIATELY.
GREEN light ‐ with YELLOW light flashing ‐ Call supplier IMMEDIATELY. Oxygen 
sensor malfunctioning; you may continue to use the concentrator. Indicators ‐ If your unit 
does not feature the O
 Sensor
2
71
Part No 1164934