Setup

The advanced setup mode lets you customize your meter.
4 types parameter are available.
P1.0: Meter configuration: (CoF)
P1.1:TDS factor(tdS)
P1.2: READY indicator:(rdy)
P1.3: ATC or Manual TC: (Atc)
P2.0: Unit :(Unt)
P2.1 select OC or OF:(t)
P2.2 select ppm or mg/L: (tdS)
P3.0: temperature parameters: (t)
P3.1: Temperature coefficient:(tCo)
P3.2: Normalization temperature:(nor)
P4.0: View calibration data (CAL)
P5.0: Electrode data:(ELE)
P7.0: Reset to factory default setting(rSt)
P1.0: Meter configuration: (CoF)
P1.1: TDS factor (tdS):
The dissolved salts increases conductivity but the effect varies from sal to
salt and is roughly linear in a given range for a given salt. The TDS
conversion factor is used to convert conductivity into TDS.
In P1.0, press ENTER to enter P1.1. Press again to see TDS factor flashes on
LCD (Fig.3). Press UP/DOWN to change the value from 0.40 to 1.00. The
default value is 0.50. Press ENTER to confirm and enter P1.2.