APPENDIX A: CONDUCTIVITY TO TDS CONVERSION FACTORS
Conductivity
TDS KCL
at 25 °C
ppm
Factor
23μS
11.6
0.5043
84μS
40.38
0.4807
447μS
225.6
0.5047
1413μS
744.7
0.527
1500μS
757.1
0.5047
2070μS
1045
0.5048
2764μS
1382
0.5
8974μS
5101
0.5685
12,880μS
7447
0.5782
15,000μS
8759
0.5839
80mS
52,168
0.6521
442: 40% sodium sulfate, 40% sodium bicarbonate and 20% sodium chloride.
GLOBAL SCIENTIFIC SUPPLY
.com
The Laboratory Supply Company!
TDS (NaCL)
TDS (442)
ppm
Factor
ppm
Factor
10.7
0.4652
14.74
0.6409
38.04
0.4529
50.5
0.6012
215.5
0.4822
300
0.6712
702.1
0.4969
1000
0.7078
737.1
0.4914
1050
0.7
1041
0.5029
1500
0.7246
1414.8
0.5119
2062.7
0.7463
4487
0.5
7608
0.8478
7230
0.5613
11,367
0.8825
8532
0.5688
13,455
0.897
48,384
0.6048
79,688
0.9961