Error code
Parameter: Conductivity
"E01", Probe is disconnected or damaged.
• Check the probe socket. If E01 still appears, replace probe.
"E02", Conductivity value is over the range limit or meter is damaged.
• Put the meter in standard solution. If E02 still appears, send back for repair.
"E03", Conductivity value is over the range limit or meter is damaged.
• Put the meter in standard solution. If E03 still appears, send back for repair.
"E04", Caused by temp. reading error.
• Refer to error code of temp. After solving the error of temp, E04 will
disappear.
"E32", Caused by IC memory error.
• send back for repair.
"E41", Caused by meter configuration error.
• Re-program the meter with correct setting.
Parameter: TDS
"E04", Caused by temp. or conductivity error.
• Refer to error code of temp. & conductivity. After solving the error of temp.
& conductivity, E04 will disappear.
Parameter: TDS
"E01", Temperature circuit is damaged.
• send back for repair.
"E02", Temp value is lower than range limit or temp. circuit is damaged.
• Put the meter in room temp. for 5 mins. If E02 still appears, send back for
repair.
"E03", Temp value is higher than range limit or temperature circuit is damaged.
• Put the meter in room temp. for 5 mins. If E03 still appears, send back for
repair.