Features

• Dual display with ATC (°C / °F switchable)
• Data hold to freeze display
• Selectable conductivity to TDS conversion factor
• USB download capabilities
• Low battery indicator.
• Auto power off
• Maximum 5 calibration points

OPERATION

Conductivity Measurement
1. Select range:
Meter is defaulted at "auto-ranging" which determines and selects the
range and gives the greatest resolution and accuracy. Alternatively, you can
manually select by pressing MODE >2 seconds. 5 ranges for selection and
"rAn" (auto ranging). It cycles from 1 to 5 and get back to auto. (Fig. 1)
Fig. 1
2. Automatic Temperature Compensation:
The meter is defaulted as ATC on. To disable ATC, refer to programming
setting P1.3 and P3.3 for manual temp. compensation.
3. ASetting correct temperature coefficient:
Factory default is 2.1% per C (temperature coefficient) and this provides
good results for most applications. Refer P3.1 if a different value is needed.
4. Select normalization temperature:
Factory default is 25 °C. Refer P3.2 if a different value is needed.