Fluke 87 V/AN Service Manual page 36

Table of Contents

Advertisement

87 V/AN
Service Manual
Step
Test Function
29
nS
Conductance
30
G
31
Diode
?
32
AC Amps
A
33
DC Amps
m ?
34
35
AC Milliamps
m[
36
DC Milliamp
37
38
?
U
39
AC Microamps
40
U
[
DC Microamps
41
42
E
43
Capacitance
44
K
45
46
Low Pass Filter
L
47
48
Peak Min/Max
49
m L
Temperature
50
51
52
Backlight
53
1. Or short test leads and use REL to offset test lead resistance.
2. Remove test leads from unit.
3. To ensure accurate measurement, the Meter and thermocouple adapter must be at the same temperature. After connecting the thermocouple
adapter to the Meter allow for reading to stabalize before recording display reading.
4. The Meter accuracy is not specified at this input signal frequency with Low-pass filter selected. The display reading shown, check that the Low-pass
filter is active and follows an expected roll-off curve.
5. Use REL to compensate for internal Meter and lead capacitance. The test leads must be disconnected from the calibrator before pushing REL.
28
Table 15. Accuracy Tests (cont.)
Range
60 nS
Open input
60 nS
100 MΩ
6 V
3.0 V dc
6 A
3.0 A, 60 Hz
6 A
3.0 A
60 mA
33 mA, 60 Hz
400 mA
330 mA, 60 Hz
60 mA
33 mA
400 mA
330 mA
600 µA
330 µA, 60 Hz
6000 µA
3300 µA, 60 Hz
600 µA
330 µA
6000 µA
3300 µA
10 nf
Open input
5
100 nf
5 nf
100 µf
9.5 µf
1000 V
400 V, 400 Hz
1000 V
400 V, 800 Hz
8 Vpp, 2 kHz Sq. Wave,
6 V
DC offset 2 V
0 °C
3
100 °C
Press backlight button
Press backlight button
Press backlight button
5500A Output
2
4
Display Reading
-0.10 to 0.10
9.80 to 10.20
2.939 to 3.061
2.968 to 3.032
2.990 to 3.010
32.65 to 33.35
326.5 to 333.5
32.89 to 33.11
329.1 to 330.9
326.5 to 333.5
3265 to 3335
328.9 to 331.1
3291 to 3309
0.21 to 0.31
4.7 to 5.3
9.2 to 9.8
372 to 408
4
226 to 340
Max = 5.896 to 6.104
Min = -1.898 to -2.102
-1.0 to 1.0
98.0 to 102.0
Backlight comes on
Backlight Intensifies
Backlight off