Performance Verification Procedure

This procedure can be used to verify the warranted characteristics of an HVD3000 or
HVD3000A probe. If the product does not meet specifications, it should be returned to a
Teledyne LeCroy service center. As there are no user accessible adjustments, there is no
adjustment procedure.

Required Equipment

The following table lists the test equipment and accessories (or their equivalents) that are
required for performance verification of the probe. This procedure is designed to minimize
the number of calibrated test instruments required. Because the input and output
connector types may vary on different brands and models of test instruments, additional
adapters or cables may be required. Only the parameters listed in boldface in the "Minimum
requirements" column must be calibrated to the accuracy indicated.
Description
Digital Multimeter
Function Generator
BNC Coaxial Cable
Calibration Fixture
Banana Plug Adapter
Insulated Banana Plug
Insulated Banana Couplers (2)
Requirements
DC: 0.1% accuracy
AC: 0.2% accuracy to measure
7 mVrms to 37 Vrms at 70 Hz
Output sine wave:
20 Vpp HVD310x and HVD3602
37 Vrms HVD3605*
* May require an amplifier to obtain
required voltage, such as Tegam
2340.
Male-to-Male 50 Ω Cable
ProBus Extender Cable
Female BNC-to-Dual
Banana Plug
Insulated BNC-to-Shrouded
Banana Plug
Operator's Manual
Example Equipment
Agilent 34401A
Fluke 8842A-09
Keithley 2001
Agilent 33120A
Stanford Research DS34
Leader LAG-120B
Pomona 5697-36
Teledyne LeCroy PROBUS-CF01
Pomona 1269 Mueller BU-00260
Mueller BU-5671-B-12-0
Mueller BU-32601-2 (R)
Mueller BU-32601-6 (B)
23