HVD3000 / HVD3000A High-Voltage Differential Probes
Checking Probe Attenuation in XStreamBrowser
Probe attenuation and other values can be found by using the XStreamBrowser application
that is installed with the oscilloscope firmware.
1. Choose File > Minimize to show the oscilloscope desktop, then double-click the
XStreamBrowser icon.
2. From the browser menu bar, choose File > Connect to Local Instrument.
3. Navigate to LeCroy.XStreamDSO > Acquisition > C1 > InputB > HVDxxxxx (where
xxxxx is your probe model number). The value is shown next to PrAttenuation.
4. Close XStreamBrowser and touch the oscilloscope icon (bottom right of desktop) to
return to the oscilloscope software.
NOTE: These steps are tailored to the Performance Verification procedure. You can find any
value at any time by changing the path to reflect the probe's actual input channel and row
(Cn > Input A|B).
26