Tài li u sưu t m c a Trung Sơn - www.tschem.com.vn - 028.3811.9991 - info@tschem.com.vn
SECTION 2,
continued
Figure 6
Sample Filtering
Tài li u sưu t m c a Trung Sơn - www.tschem.com.vn - 028.3811.9991 - info@tschem.com.vn
After dilution and measurement, calculate the actual result as follows:
1. Calculate the dilution factor:
Total Volume
Dilution Factor
=
----------------------------------------- -
Sample Volume
Where total volume = sample + dilution water
Example: 20 mL of sample + 80 mL of dilution water = 100 mL total
100
Dilution Factor
=
--------- -
20
2. Calculate the Final Turbidity Value:
×
Measured Results Dilution Factor
For example, if the measured turbidity value is 1100 NTU, the final turbidity value is
calculated as:
×
1100
5
5500
=
30
=
5
=
Actual NTU